VARiON
®
Plus
700 IQ H
Commissioning
25
ba76207e02
12/2020
If the setting
VARiON Cl
hide
was additionally selected:
Calib. history Cl
Do not download
Transmit to log book
The selection of
Transmit to log book
gen-
erates a log book message with the cali-
bration history of the potassium electrode.
When opening the setting table again the
setting is reset to
Do not download
.
Chloride compens.
Automatic
When a chloride electrode is mounted, the
chloride compensation takes places auto-
matically only. The value of the chloride
concentration measured at the time the
setting table was opened is displayed in
the next line (
Chloride conc.
).
Chloride conc.
Display of the measured chloride value
(
0.1 ... 1000 mg/l)
.
Temperature mode
(only with
Measuring
mode
:
NO3-N
or
NO3
)
°C
°F
Unit of the measured temperature value
(Celsius, Fahrenheit).
Temp. adjustment
(only with
Measuring
mode
:
NO3-N
or
NO3
)
-1.5 °C ... +1.5 °C
The temperature compensation function
enables the temperature sensor to be
balanced against a reference temperature
measurement (displacement of the zero
point by ±1.5 °C).
Notes:
Due to the thermal capacity of the sen-
sor, it is necessary to place it in a con-
tainer with at least 2 liters of water.
Leave the sensor in this container for at
least 15 minutes while stirring occa-
sionally, then carry out the adjustment.
If the temperature difference of the
water and sensor is > 10°C, leave the
sensor in the container for at least one
hour while stirring occasionally.
Conc. offset
-1.0 mg/l ... +1.0 mg/l
Constant value that is added to the mea-
sured value.
Save and quit
The system confirms the saving of the set-
tings and the display switches to the next
higher level.
Quit
The display switches to the next higher
level without storing the new settings.
Menu item
Selection/Values
Explanations