38
Calibration
ODO mg/L Calibration
1.
Place the ODO and conductivity/temperature sensor into a water
sample that has been titrated by the Winkler method to determine
the dissolved oxygen concentration in mg/L.
2.
Push the
Cal
key, then select
ODO
. Select
DO mg/L
.
3.
Select
Calibration value.
4.
Enter the dissolved oxygen concentration of the sample in mg/L.
5.
Observe the actual measurement readings for stability (white line
on graph shows no significant change for 40 seconds), then select
Accept Calibration
(
Figure 49
). “Calibration successful!” will be
displayed in the message area.
6.
Rinse the bulkhead and sensors in clean water then dry.
ODO Zero Point Calibration
1.
Place the ODO and Conductivity/Temperature sensors in a
solution of zero DO.
NOTE:
A zero DO solution can be made by dissolving
approximately 8-10 grams of sodium sulfite into 500 mL
of tap water. Mix the solution thoroughly. It may take the
solution 60 minutes to be oxygen-free.
2.
Push the
Cal
key, then select
ODO
. Select
Zero
.
3.
Observe the actual measurement readings for stability (white line
on graph shows no significant change for 40 seconds), then select
Accept Calibration
(
Figure 50
). “Calibration successful!” will be
displayed in the message area.
4.
Thoroughly rinse the bulkhead and sensors in clean water then
dry.
5.
Perform a ODO % water-saturated air calibration after performing
a zero point calibration.
Figure 49
Calibrate ODO mg/L
Figure 50
Calibrate ODO zero point
Summary of Contents for YSI ProDSS
Page 1: ...USER MANUAL ProDSS USER MANUAL DOCUMENT 626973 01REF...
Page 2: ...i This page left intentionally blank...
Page 58: ...56 Calibration This page left intentionally blank...
Page 74: ...72 Maintenance and Storage This page left intentionally blank...
Page 90: ...88 Appendix A DO Calibration Values This page left intentionally blank...