background image

6

7

CONNECTING THE PROBE/CABLE ASSEMBLY TO THE 
INSTRUMENT

To connect the cable, align the keys on the cable connector to the slots 
on the instrument connector. Push together firmly and then twist the 
outer ring until it locks into place (figure 4). This connection is water-
proof.

Figure 4, Note the keyed connector.  

MEMBRANE INSTALLATION

The sensing end of the dissolved oxygen sensor is shipped with a 
protective cap that needs to be removed before using. Additionally, it 
is very important to install a new membrane with electrolyte solution 
onto the sensor after removing the cap.  

Prepare the electrolyte solution according to the instructions on the 
bottle. After mixing, allow the solution to sit for 1 hour. This will help 
prevent air bubbles from later developing under the membrane. 
Ensure you are using the correct electrolyte solution for your sensor. 
Galvanic sensors utilize electrolyte with a light blue label on the bottle 
and polarographic sensors utilize electrolyte with a white label on the 
bottle. The dissolved oxygen sensor is supplied with cap membranes 
specific to the sensor type ordered (polarographic or galvanic). 5913 
and 5914 membrane kits are for Galvanic sensors and the 5908 and 
5909 membrane kits are for Polarographic sensors.

Remove and discard or save the protective cap from the dissolved 
oxygen sensor by pulling it straight off. Thoroughly rinse the sensor 
tip with distilled or deionized water. Fill the cap membrane 3⁄4 full 
of electrolyte solution, then tap the cap with a finger to release any 
trapped air. Be careful not to touch the membrane portion of the cap. 
Thread the membrane cap onto the sensor, moderately tight. Do not 
use a tool. It’s typical for some of the electrolyte solution to spill over. It 
is best to allow the new cap to remain on the sensor overnight before 

 

calibrating. For detailed instructions on changing a cap membrane, 
see the

 

Care, Maintenance and Storage section of this manual.  

RUN SCREEN

Press the power/backlight key 

to turn the instrument on. The 

instrument will run through a self test and briefly display a splash screen 
with system information before displaying the main Run screen (figure 
5). A language, dissolved oxygen sensor and membrane selection 
menu will display the first time the Pro1020 is powered on. See the 
First Power On section of this manual for more information. 

Figure 5, Run screen.

BACKLIGHT

Once the instrument is powered on, pressing the power/backlight key 

will turn on the display backlight. The backlight will remain on until 

the key is pressed again or after two minutes of not pressing any key 
on the keypad.  

POWERING OFF

To turn the instrument off, press and hold the power/backlight key  
for three seconds.  

NAVIGATION

The up   and down   arrow keys allow you to navigate through the 
functions of the Pro1020.

Summary of Contents for YSI Pro1020

Page 1: ...USER MANUAL English Fran ais Espa ol Portugu s Pro1020 User Manual DOCUMENT 605187...

Page 2: ...1 Battery Installation 1 Key Pad 2 Connecting the Sensor and Cable 3 Membrane Installation 6 Run Screen 7 Backlight 7 Powering Off 7 Navigation 7 First Power On 8 System Setup Menu 9 DO Local 10 Last...

Page 3: ...YSI will repair or replace at its sole discretion free of charge any product that YSI determines to be covered by this warranty To exercise this warranty call your local YSI representative or contact...

Page 4: ...arts or materials are damaged or missing contact YSI Customer Service at 800 897 4151 1 937 767 7241 or the authorized YSI distributor from whom the instrument was purchased BATTERY INSTALLATION The i...

Page 5: ...he bottom of the Run screen and to increase numerical inputs 3 Power and Backlight Press once to turn instrument on Press a second time to turn backlight on Press a third time to turn backlight off Pr...

Page 6: ...Next remove the o ring and slide the lid off the sensor INSTALLING THE DISSOLVED OXYGEN SENSOR Before installing a sensor or connecting the cable to the instrument the Sensor Type must be configured f...

Page 7: ...cap from the dissolved oxygen sensor by pulling it straight off Thoroughly rinse the sensor tip with distilled or deionized water Fill the cap membrane 3 4 full of electrolyte solution then tap the ca...

Page 8: ...dissolved oxygen sensor type and membrane type Use the up or down arrow keys to highlight the appropriate language sensor and membrane then press enter to confirm figure 6 The Sensor Type must be con...

Page 9: ...instead of during calibration Example Instrument 1 with DO Local enabled At 737 mmHg barometric pressure roughly 841 ft above sea level the instrument would calibrate to 100 When taking measurements...

Page 10: ...ompatible sensors for use with a field cable Polarographic This sensor has a black sensor body and is engraved with the model number 2003 Polarographic will be abbreviated Polaro in the instrument men...

Page 11: ...lay steadily and the instrument will beep twice if Audio is turned on DO Auto Stable can be set to a change of 0 0 to 1 9 over 3 to 19 seconds ISE Auto Stable can be set to a change of 0 0 to 9 9 over...

Page 12: ...nge the language Highlight the desired language English Spanish Portuguese or French and press enter to enable The enabled language will have an X in the box next to it Highlight the ESC Exit box and...

Page 13: ...ate formula for this uncorrection is True BP Corrected BP 2 5 Local Altitude in ft above sea level 100 Although the Pro1020 barometer range is 400 0 to 999 9 mmHg you can not adjust the value across t...

Page 14: ...ubleshooting guide for possible solutions CALIBRATING IN PERCENT DO LOCAL ENABLED Perform this calibration procedure when DO Local is enabled and Quick DO Cal is disabled in the System Setup menu 1 Pe...

Page 15: ...5 If necessary use the up and down arrow keys to adjust the pH buffer value Note the pH mV reading which ideally should be between 50 and 50 in buffer 7 6 Press enter to continue to second point 7 Ri...

Page 16: ...membrane If using the 1001A amplified pH sensor be sure to use the guard extender supplied in the 605216 amplified pH sensor kit The extender is screwed onto the sensor bulkhead before the guard or c...

Page 17: ...enter to regain access to scrolling through the saved data The data set displayed is indicated by the data set number 01 through 50 ERASING DATA While viewing saved data press the enter key to access...

Page 18: ...or connector Check the o ring and the groove for any excess grease or contamination If contamination is evident clean the o ring and nearby plastic parts with lens cleaning tissue or equivalent lint f...

Page 19: ...the sensor in clean water wipe with a cotton swab saturated with clean water not DI water and then rerinse with clean water To be certain that all traces of the acid are removed i from the sensor cre...

Page 20: ...d in the membrane kit Do not sand the electrodes every membrane change as this is not routine maintenance In fact visually the anode may appear tarnished and operate properly YSI recommends using the...

Page 21: ...embrane should be changed if a bubbles are visible under the membrane b dried electrolyte is visible around the membrane and c if the sensor shows unstable readings or other sensor related symptoms Th...

Page 22: ...em for service Symptom Possible Solution Instrument will not calibrate pH or ORP instrument displays Calibration Over Calibration Under or Unstable Reading during calibration 1 Verify correct ISE sens...

Page 23: ...specifications represent typical performance and are subject to change without notice For the latest product specification information please visit YSI s website at ysi com or contact YSI Tech Suppor...

Page 24: ...E blue cap 6 each includes electrolyte Used with Galvanic sensor All cables include a temperature sensor The pH or ORP and Dissolved Oxygen sensors are sold separately D E C L A R AT I O N O F C O N F...

Page 25: ...country and region and users are expected to understand and follow the battery disposal requirements for their specific locale C O N TA C T I N F O R M AT I O N ORDERING AND TECHNICAL SUPPORT Telephon...

Page 26: ...Item 605187 Rev A January 2014 2014 YSI Incorporated...

Reviews: