14

15

Selecting a Dissolved Oxygen Membrane:

Membrane 

Type

Flow Dependence 

After 4 Minutes

Required Sample 

Movement

Typical Response 

Time (T-95)

5913, 5908 

Yellow

25%

6 inches/second

8 seconds

5914, 
5909 Blue

18%

3 inches/second

17 seconds

DO UNITS

Highlight DO Units and press enter to open a submenu that allows you 
to select the dissolved oxygen units to be displayed on the Run screen. 
Highlight a unit and press enter to enable or disable it. An enabled 
dissolved oxygen unit will have an ‘X’ in the box next to it. Highlight the 
ESC-Exit box along the bottom of the display and press enter to save 
any changes and to close the DO units submenu. 

There are three options for displaying dissolved oxygen:  

• 

mg/L will show DO readings in milligrams per liter on a scale 
from 0 to 50 mg/L.

• 

ppm (parts per million) is equivalent to mg/L and will show the 
DO reading on a scale from 0 to 50 ppm.  

• 

% will show DO readings in a % saturation from 0 to 500%. This 
value will be expressed as %L when DO Local% is enabled. 

mg/L and ppm cannot be enabled and therefore displayed at the 
same time.  DO% and mg/L 

or

 ppm can be enabled and displayed 

simultaneously.   

AUTO STABLE

Auto Stable utilizes preset values to indicate when a reading is stable. 
The preset values are adjustable in the System Setup menu. The user 
can input a % change in measurement reading over ‘x’ amount of time 
in seconds (3 - 19). There are two separate Auto Stable controls, one for 
dissolved oxygen readings (DO Auto Stable) and one for ISE readings 
(ISE Auto Stable). They are located on the second page of the System 
Setup menu.

When Auto Stable is enabled, an

AS

 symbol will display next to the 

reading on the Run screen and blink during stabilization. When the 
ISE and/or conductivity reading stabilizes based on the Auto Stable 

settings, the

AS

 

symbol will display steadily and the instrument will 

beep twice if Audio is turned on. 

DO Auto Stable can be set to a % change of 0.0 to 1.9% over 3 to 19 
seconds.  

ISE Auto Stable can be set to a % change of 0.0 to 9.9% over 3 to 19 
seconds. The auto stable criteria is applied to the pH measurement or 
the ORP mV reading depending on which sensor is enabled in the ISE 
Sensor menu. 

To enable Auto Stable, highlight either DO Auto Stable or ISE Auto 
Stable, then press enter to open the submenu. Next, use the up or 
down arrow keys to highlight the % change or seconds (secs) input 
field, then press enter to make the highlighted field adjustable. Use the 
up or down arrow keys to adjust the selected value, then press enter to 
confirm changes. Once you have confirmed any changes, highlight the 
ESC-Exit box along the bottom of the display and press enter to close 
the Auto Stable submenu. To disable Auto Stable, set the % Change 
input to 0.0. 

ISE SENSOR TYPE

ISE Sensor Type sets the type of ISE sensor being used; either pH  
(model #1001 or #1001A) or ORP (model #1002).

Use the up or down arrow keys to highlight ISE Sensor Type, then press 
enter to open a submenu. Highlight the sensor type corresponding 
to the sensor installed on the cable and press enter to confirm. The 
enabled sensor type will have an ‘X’ in the box next to it. Next, use the 
down arrow key to highlight the ESC – Exit box, then press enter to 
save changes and to close the sensor submenu.  

ISE UNITS

Highlight ISE Units and press enter to open a submenu that allows you 
to select the ISE units to be displayed on the Run screen. Highlight a 
unit and press enter to enable or disable it. An enabled ISE unit will 
have an ‘X’ in the box next to it. Highlight the ESC-Exit box along the 
bottom of the display and press enter to save any changes and to close 
the ISE Units submenu. 

When pH is enabled in the ISE Sensor Type menu, there are two 
selectable measurement units: pH and pH mV.  pH mV is the sensor’s 
electrical measurement signal before being converted into pH units. pH 

Summary of Contents for YSI Pro1020

Page 1: ...USER MANUAL English Fran ais Espa ol Portugu s Pro1020 User Manual DOCUMENT 605187...

Page 2: ...1 Battery Installation 1 Key Pad 2 Connecting the Sensor and Cable 3 Membrane Installation 6 Run Screen 7 Backlight 7 Powering Off 7 Navigation 7 First Power On 8 System Setup Menu 9 DO Local 10 Last...

Page 3: ...YSI will repair or replace at its sole discretion free of charge any product that YSI determines to be covered by this warranty To exercise this warranty call your local YSI representative or contact...

Page 4: ...arts or materials are damaged or missing contact YSI Customer Service at 800 897 4151 1 937 767 7241 or the authorized YSI distributor from whom the instrument was purchased BATTERY INSTALLATION The i...

Page 5: ...he bottom of the Run screen and to increase numerical inputs 3 Power and Backlight Press once to turn instrument on Press a second time to turn backlight on Press a third time to turn backlight off Pr...

Page 6: ...Next remove the o ring and slide the lid off the sensor INSTALLING THE DISSOLVED OXYGEN SENSOR Before installing a sensor or connecting the cable to the instrument the Sensor Type must be configured f...

Page 7: ...cap from the dissolved oxygen sensor by pulling it straight off Thoroughly rinse the sensor tip with distilled or deionized water Fill the cap membrane 3 4 full of electrolyte solution then tap the ca...

Page 8: ...dissolved oxygen sensor type and membrane type Use the up or down arrow keys to highlight the appropriate language sensor and membrane then press enter to confirm figure 6 The Sensor Type must be con...

Page 9: ...instead of during calibration Example Instrument 1 with DO Local enabled At 737 mmHg barometric pressure roughly 841 ft above sea level the instrument would calibrate to 100 When taking measurements...

Page 10: ...ompatible sensors for use with a field cable Polarographic This sensor has a black sensor body and is engraved with the model number 2003 Polarographic will be abbreviated Polaro in the instrument men...

Page 11: ...lay steadily and the instrument will beep twice if Audio is turned on DO Auto Stable can be set to a change of 0 0 to 1 9 over 3 to 19 seconds ISE Auto Stable can be set to a change of 0 0 to 9 9 over...

Page 12: ...nge the language Highlight the desired language English Spanish Portuguese or French and press enter to enable The enabled language will have an X in the box next to it Highlight the ESC Exit box and...

Page 13: ...ate formula for this uncorrection is True BP Corrected BP 2 5 Local Altitude in ft above sea level 100 Although the Pro1020 barometer range is 400 0 to 999 9 mmHg you can not adjust the value across t...

Page 14: ...ubleshooting guide for possible solutions CALIBRATING IN PERCENT DO LOCAL ENABLED Perform this calibration procedure when DO Local is enabled and Quick DO Cal is disabled in the System Setup menu 1 Pe...

Page 15: ...5 If necessary use the up and down arrow keys to adjust the pH buffer value Note the pH mV reading which ideally should be between 50 and 50 in buffer 7 6 Press enter to continue to second point 7 Ri...

Page 16: ...membrane If using the 1001A amplified pH sensor be sure to use the guard extender supplied in the 605216 amplified pH sensor kit The extender is screwed onto the sensor bulkhead before the guard or c...

Page 17: ...enter to regain access to scrolling through the saved data The data set displayed is indicated by the data set number 01 through 50 ERASING DATA While viewing saved data press the enter key to access...

Page 18: ...or connector Check the o ring and the groove for any excess grease or contamination If contamination is evident clean the o ring and nearby plastic parts with lens cleaning tissue or equivalent lint f...

Page 19: ...the sensor in clean water wipe with a cotton swab saturated with clean water not DI water and then rerinse with clean water To be certain that all traces of the acid are removed i from the sensor cre...

Page 20: ...d in the membrane kit Do not sand the electrodes every membrane change as this is not routine maintenance In fact visually the anode may appear tarnished and operate properly YSI recommends using the...

Page 21: ...embrane should be changed if a bubbles are visible under the membrane b dried electrolyte is visible around the membrane and c if the sensor shows unstable readings or other sensor related symptoms Th...

Page 22: ...em for service Symptom Possible Solution Instrument will not calibrate pH or ORP instrument displays Calibration Over Calibration Under or Unstable Reading during calibration 1 Verify correct ISE sens...

Page 23: ...specifications represent typical performance and are subject to change without notice For the latest product specification information please visit YSI s website at ysi com or contact YSI Tech Suppor...

Page 24: ...E blue cap 6 each includes electrolyte Used with Galvanic sensor All cables include a temperature sensor The pH or ORP and Dissolved Oxygen sensors are sold separately D E C L A R AT I O N O F C O N F...

Page 25: ...country and region and users are expected to understand and follow the battery disposal requirements for their specific locale C O N TA C T I N F O R M AT I O N ORDERING AND TECHNICAL SUPPORT Telephon...

Page 26: ...Item 605187 Rev A January 2014 2014 YSI Incorporated...

Reviews: