TruLab pH 1310
Data storage
ba76156e01
03/2014
49
Settings
You can configure the automatic storage function with the following settings:
Starting the auto-
matic storage func-
tion
To start the automatic storage function, select
Continue
with
<
><
>
and
confirm with
<MENU/ENTER>
. The meter switches to the measured value dis-
play.
The active automatic storage function can be recognized by the progress bar
in the function display. The progress bar indicates the remaining storage dura-
tion.
Energy saving mode
([Eco mode])
If the automatic storing function is active, the meter provides an energy saving
mode ([Eco mode]) to avoid unnecessary energy consumption. The energy
Menu item
Possible set-
ting
Description
ID number
1 ... 10000
ID number for the dataset series.
Interval
1 s, 5 s, 10 s,
30 s, 1 min,
5 min,
10 min,
15 min,
30 min,
60 min
Storage interval.
The lower limit of the storage
interval can be restricted by the
number of free storage locations.
The upper limit is restricted by the
storage duration.
Duration
1 min ... x min
Storage duration.
Specifies after which time the
automatic storage should be ter-
minated.
The lower limit of the storage
duration is restricted by the stor-
age interval.
The upper limit is restricted by the
number of free storage locations.
1 Remaining storage duration
2 Graphical display of the storage duration
If the automatic storage function is activated, only the following
keys are active: Softkeys,
<M>
,
<STO__>
and
<On/Off>
. The
other keys and the automatic switch-off function are deactivated.
1
2
pH
6.949
25.0
°C
0d03h00min
01.02.2014
08:00
0
14
USB output
Eco mode
Summary of Contents for YSI pH 1310P
Page 70: ...Index TruLab pH 1310 70 ba76156e01 03 2014...
Page 72: ...Contact Information TruLab pH 1310 72 ba76156e01 03 2014...
Page 73: ......