MultiLab 4010-1
pH value
ba76141e03
03/2014
35
Continuing with
three- to five-point
calibration
14. Wait for the measurement with stability control to be completed or ter-
minate the stability control and take over the calibration value with
<MENU/ENTER>
.
The pH value of the buffer solution is displayed.
15. Set the nominal buffer value for the measured temperature with
<
><
>
.
16. Accept the calibration value with
<MENU/ENTER>
.
The calibration display for the next buffer appears (voltage display).
17. If necessary, finish the calibration procedure as a two-point calibration
with
<M>
.
The calibration record is displayed.
pH
4.035
24.8
°C
ConCal
2
01.07.2011
08:00
Buffer
18. Thoroughly rinse the sensor with deionized water.
19. Immerse the sensor in the next buffer solution.
20. Start the measurement with
<MENU/ENTER>
.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter
flashes.
21. Wait for the measurement with stability control to be completed or ter-
minate the stability control and take over the calibration value with
<MENU/ENTER>
.
The pH value of the buffer solution is displayed.
pH
9.958
24.8
°C
ConCal
3
01.07.2011
08:00
Buffer