Maintenance and cleaning

SenTix

®

ORP-T 900(-P)

8

ba75845e03

02/2017

5

Maintenance and cleaning

During operation, a small amount of reference electrolyte leaks through the 
junction from the electrode into the test sample. If the level of reference elec-
trolyte becomes too low with time, refill it through the refilling opening.

Refilling the

reference

electrolyte

Refilling is very easy using a dropping bottle. Proceed as follows:

Cut off the tip of the dropping bottle at a right angle until the opening in the 
tip can be seen

Open the refilling opening of the electrode

Press the tip of the dropping bottle into the refilling opening while turning 
it slightly

Pump several small quantities of the reference electrolyte into the stem 
using the dropper bottle

Pull the dropping bottle out of the refilling opening while turning it slightly 
as necessary.

Cleaning

Remove water-soluble contamination by rinsing with deionized water. 
Remove other contamination as follows:

After cleaning

Rinse the electrode with deionized water.

Contamination

Cleaning procedure

Fat and oil

Rinse with water containing household wash-
ing-up liquid

Lime and hydroxide deposits Rinse with citric acid (10 % by weight)

Protein

Immerse in pepsin cleaning solution PEP/pH 
for approx. 1 hour.
Note: Make sure the level of the reference 
electrolyte is above that of the cleaning solu-
tion.

Summary of Contents for WTW SenTix ORP-T 900

Page 1: ...ba75845e03 02 2017 SenTix ORP T 900 P ORP ELECTRODE SenTix ORP T 900 P OPERATING MANUAL For Sales Service Contact 2650 E 40th Ave Denver CO 80205 Phone 303 320 4764 Fax 303 322 7242 1 800 833 7958 ww...

Page 2: ...2 ba75845e03 02 2017 SenTix ORP T 900 Copyright 2017 Xylem Analytics Germany GmbH Printed in Germany...

Page 3: ...is connected and is used for measurement and for measured value documentation Storing the calibration data in the sensor ensures that the correct slope and asymmetry are automatically used if the sen...

Page 4: ...and index finger to press the clips of the connector together and pull the connector out of the plug step 2 Closing the plug connection Make sure that the plug connection is completely dry and clean...

Page 5: ...l of the meter and observe the following rules while doing so Preparing the sensor for measurement 2 3 General rules for measuring Make sure the refilling opening for the reference electrolyte solutio...

Page 6: ...orrect The junction must be com pletely submersed in the solution The junction is in the area of the bottom end of the shaft see arrow At the same time the level of the reference electrolyte must be a...

Page 7: ...horten the lifetime of the electrode These are Strong acids or lyes hydrofluoric acid organic solvents oils fats bro mides sulfides iodides proteins High temperatures High changes in pH and temperatur...

Page 8: ...e tip of the dropping bottle into the refilling opening while turning it slightly Pump several small quantities of the reference electrolyte into the stem using the dropper bottle Pull the dropping bo...

Page 9: ...value and the ORP voltage UH of the test solution where the latter must be estimated for the first measurement The type of pretreatment can then be determined using the following diagram where UH is b...

Page 10: ...Platinum circle Connection cable Lengths SenTix ORP T 900 1 5 m SenTix ORP T 900 P 1 5 3 6 10 15 25 40 60 100 m Diameter 4 3 mm Smallest allowed bend radius Fixed installation 20 mm Flexible use 60 mm...

Page 11: ...ion 3 bottles of 250 ml each PEP pH 109 648 Connection cableSenTix ORP T 900 P meter Description Model Order no IDS connection cable 1 5 m AS IDS 1 5 903 850 IDS connection cable 3 m AS IDS 3 903 851...

Page 12: ...Disposal SenTix ORP T 900 P 12 ba75845e03 02 2017...

Page 13: ......

Page 14: ...buildings factories and farms In more than 150 countries we have strong long standing relationships with customers who know us for our powerful combination of leading product brands and applications...

Reviews: