Maintenance and cleaning
SenTix
®
ORP-T 900(-P)
8
ba75845e03
02/2017
5
Maintenance and cleaning
During operation, a small amount of reference electrolyte leaks through the
junction from the electrode into the test sample. If the level of reference elec-
trolyte becomes too low with time, refill it through the refilling opening.
Refilling the
reference
electrolyte
Refilling is very easy using a dropping bottle. Proceed as follows:
Cut off the tip of the dropping bottle at a right angle until the opening in the
tip can be seen
Open the refilling opening of the electrode
Press the tip of the dropping bottle into the refilling opening while turning
it slightly
Pump several small quantities of the reference electrolyte into the stem
using the dropper bottle
Pull the dropping bottle out of the refilling opening while turning it slightly
as necessary.
Cleaning
Remove water-soluble contamination by rinsing with deionized water.
Remove other contamination as follows:
After cleaning
Rinse the electrode with deionized water.
Contamination
Cleaning procedure
Fat and oil
Rinse with water containing household wash-
ing-up liquid
Lime and hydroxide deposits Rinse with citric acid (10 % by weight)
Protein
Immerse in pepsin cleaning solution PEP/pH
for approx. 1 hour.
Note: Make sure the level of the reference
electrolyte is above that of the cleaning solu-
tion.