Xylem WTW SenTix ORP-T 900 Operation Manual Download Page 3

SenTix

®

ORP-T 900(-P)

General information

3

ba75845e03

02/2017

1

General information

Automatic sensor

recognition

The sensor electronics with the stored sensor data are in the connecting 
head of the electrode. The data include, among other things, the sensor type 
and series number. In addition, the calibration data are stored in the sensor 
with each calibration and the calibration history is recorded (the last 10 cali-
brations). The data is recalled by the meter when the sensor is connected 
and is used for measurement and for measured value documentation. 

Storing the calibration data in the sensor ensures that the correct slope and 
asymmetry are automatically used if the sensor is operated with different 
meters. On the other hand, different calibrated sensors can be used with one 
meter without the need to recalibrate.

The digital transmission technique guarantees the failure-free communica-
tion with the meter even with long connection cables. The sensor firmware  
can be updated via the meter.

2

Commissioning, measuring, checking

For ORP electrodes with platinum electrode, please follow the 
instructions in section 6 A

CTIVATING

 

PLATINUM

 

ELECTRODES

.

Summary of Contents for WTW SenTix ORP-T 900

Page 1: ...ba75845e03 02 2017 SenTix ORP T 900 P ORP ELECTRODE SenTix ORP T 900 P OPERATING MANUAL For Sales Service Contact 2650 E 40th Ave Denver CO 80205 Phone 303 320 4764 Fax 303 322 7242 1 800 833 7958 ww...

Page 2: ...2 ba75845e03 02 2017 SenTix ORP T 900 Copyright 2017 Xylem Analytics Germany GmbH Printed in Germany...

Page 3: ...is connected and is used for measurement and for measured value documentation Storing the calibration data in the sensor ensures that the correct slope and asymmetry are automatically used if the sen...

Page 4: ...and index finger to press the clips of the connector together and pull the connector out of the plug step 2 Closing the plug connection Make sure that the plug connection is completely dry and clean...

Page 5: ...l of the meter and observe the following rules while doing so Preparing the sensor for measurement 2 3 General rules for measuring Make sure the refilling opening for the reference electrolyte solutio...

Page 6: ...orrect The junction must be com pletely submersed in the solution The junction is in the area of the bottom end of the shaft see arrow At the same time the level of the reference electrolyte must be a...

Page 7: ...horten the lifetime of the electrode These are Strong acids or lyes hydrofluoric acid organic solvents oils fats bro mides sulfides iodides proteins High temperatures High changes in pH and temperatur...

Page 8: ...e tip of the dropping bottle into the refilling opening while turning it slightly Pump several small quantities of the reference electrolyte into the stem using the dropper bottle Pull the dropping bo...

Page 9: ...value and the ORP voltage UH of the test solution where the latter must be estimated for the first measurement The type of pretreatment can then be determined using the following diagram where UH is b...

Page 10: ...Platinum circle Connection cable Lengths SenTix ORP T 900 1 5 m SenTix ORP T 900 P 1 5 3 6 10 15 25 40 60 100 m Diameter 4 3 mm Smallest allowed bend radius Fixed installation 20 mm Flexible use 60 mm...

Page 11: ...ion 3 bottles of 250 ml each PEP pH 109 648 Connection cableSenTix ORP T 900 P meter Description Model Order no IDS connection cable 1 5 m AS IDS 1 5 903 850 IDS connection cable 3 m AS IDS 3 903 851...

Page 12: ...Disposal SenTix ORP T 900 P 12 ba75845e03 02 2017...

Page 13: ......

Page 14: ...buildings factories and farms In more than 150 countries we have strong long standing relationships with customers who know us for our powerful combination of leading product brands and applications...

Reviews: