Technical data
Oxi 3205
36
ba75797e05
12/2017
7.2
Measuring ranges, resolution, accuracy
Note
The accuracy values specified here apply exclusively to the meter. The
accuracy of the D.O. sensors has also to be taken into account.
Measuring ranges,
resolution
Note: The values quoted in brackets apply especially for the
DurOx 325 sensor.
Parameter
Measuring range
Resolution
D.O. concentration
[mg/l]
0 ... 20.00 (0 ... 20.0)
0 ... 90.0 (0 ... 90)
0.01 (0.1)
0.1 (1)
Saturation [%]
0 ... 200.0 (0 ... 200)
0 ... 600
0.1 (1)
1
D.O. partial pressure
[mbar]
0 ... 200.0 (0 ... 200)
0 ... 1250
0.1 (1)
1
T [°C]
0 ... 50.0
0.1
Accuracy (± 1 digit)
Parameter
Accuracy
D.O. concentration
[mg/l]
± 0.5 % of measured value
at ambient tempe 5 °C ... + 30 °C
Saturation [%]
± 0.5% of measured value
when measuring in the range of ± 10 K
around the calibration temperature
D.O. partial pressure
[mbar]
± 0.5 % of measured value
at ambient tempe 5 °C ... + 30 °C
T [°C]
/ temperature sensor
NTC 30
± 0,1
PT 1000
± 0,1
Correction functions
Temperature
compensation
Accuracy better than 2 % at 0 ... + 40 °C
Salinity correction
35 SAL
Air pressure
correction
Automatic through integrated pressure
sensor in the range of 500 ... 1100 mbar
Summary of Contents for WTW Oxi 3205
Page 2: ...2 ba75797e05 12 2017 Oxi 3205 Copyright 2017 Xylem Analytics Germany GmbH Printed in Germany ...
Page 12: ...Safety Oxi 3205 12 ba75797e05 12 2017 ...
Page 30: ...Operation Oxi 3205 30 ba75797e05 12 2017 ...
Page 34: ...What to do if Oxi 3205 34 ba75797e05 12 2017 ...
Page 38: ...Technical data Oxi 3205 38 ba75797e05 12 2017 ...
Page 42: ...Lists Oxi 3205 42 ba75797d05 12 2017 ...
Page 43: ......