Operation
Oxi 3205
24
ba75797e05
12/2017
4.4.3
Settings for D.O. sensors
(menu or measurement and calibration settings)
The settings are available in the menu for measurement and calibration
settings. To open the settings, display the required parameter in the
measured value display and press the
<ENTER>
or
<ENTER>
key.
After completing the settings, switch to the measured value display with
<M>
.
Menu item
Possible
setting
Description
Calibration record
-
Displays the calibration
record of the last calibration.
Calibration interval
1 ... 999 d
Calibration interval
for the
D.O. sensor (in days).
The meter reminds you to
calibrate regularly by the
flashing sensor symbol in
the measuring window.
Temperature unit
°C
°F
Temperature unit,
degrees Celsius or degrees
Fahrenheit.
All temperatures are
displayed with the selected
unit.
Sal correction
On
Off
Manual salt content
correction for concentration
measurements.
Reset
-
Resets all sensor settings to
the delivery condition (see
section 4.5.1).