NitraVis 70x IQ
Technical data
43
ba75951e03
05/2017
8
Technical data
8.1
Measurement characteristics
Measuring principle Spectrophotometric Absorption measurement; integrated microprocessor elec
-
tronics, shielded 2-wire connection for power and data transmission.
Light source
Detector
Wavelengths range 200 - 720 nm
Measuring gap
(optical layer
thickness)
Nitrate
measurement
1 mm gap width
Nitrate
measurement,
5 mm gap width
Lamp type
Xenon flashlamp
Photo diodes
256
NitraVis 701 IQ
1 mm
NitraVis 705 IQ
5 mm
Measuring
location
Measuring mode
Measuring range
Resolution
Inlet
NO3
0.0 ... 300.0 mg/L
0.1 mg/L
NO3-N
0.00 ... 60.00 mg/L
0.01 mg/L
Activation
NO3
0.0 ... 300.0 mg/L
0.1 mg/L
NO3-N
0.00 ... 60.00 mg/L
0.01 mg/L
Outlet
NO3
0.0 ... 750.0 mg/L
0.1 mg/L
NO3-N
0.0 ... 150.0 mg/l
0.1 mg/L
Measuring
location
Measuring mode
Measuring range
Resolution
Outlet
NO3
0.0 ... 250.0 mg/L
0.1 mg/L
NO3-N
0.00 ... 50.00 mg/L
0.01 mg/L
The quoted measuring ranges are nominal measuring ranges that
are theoretically possible. In practice, real measuring ranges exist
that are given by the limits of photometric determination. The limits
are significantly influenced by the light scattering due to solids and
the absorption of accompanying substances (sample matrix).
Therefore, no fixed measuring range can be determined before
-
hand.