Commissioning
NitraVis 70x IQ
14
ba75951e03
05/2017
3.3.3 Connecting the sensor to the IQ S
ENSOR
N
ET
Connection cable A sensor connection cable of the SACIQ or SACIQ SW type is required to
connect the sensor. The cable is available in different lengths. Compared to the
standard model SACIQ, the SACIQ SW sensor connection cable is optimized
regarding its corrosion resistance in seawater and brackish water and adapted
for use in conjunction with the NitraVis 70x IQ. Information on this and other
IQ S
ENSOR
N
ET
accessories is given in the WTW catalog and on the Internet.
Are the plug
connections dry?
Before connecting the sensor and sensor connection cable, please make sure
the plug connections are dry. If moisture gets into the plug connections, first dry
the plug connections (dab them dry or blow them dry using compressed air).
NOTE
Do not suspend the sensor on the sensor connection cable but use the suitable
mounting equipment. Information on this and other IQ S
ENSOR
N
ET
accessories
is given in the WTW
catalog and on the Internet.
How to connect the SACIQ (SW) sensor connection cable to the ter
-
minal strip of an MIQ module is described in chapter 3 I
NSTALLATION
of the IQ S
ENSOR
N
ET
system operating manual.