pH value
Multi 9430
34
ba75966e04
04/2015
Continuing with two-
point calibration
9.
Set the nominal buffer value for the measured temperature with
<
><
>
.
10. Accept the calibration value with
<ENTER>
.
The calibration display for the next buffer appears (voltage display).
11. If necessary, finish the calibration procedure as a single-point calibration
with
<M>
.
The calibration record is displayed.
For
single-point calibration
, the instrument uses the Nernst slope
(-59.2 mV/pH at 25 °C) and determines the zero point of the IDS pH
sensor.
12. Thoroughly rinse the sensor with deionized water.
13. Immerse the pH sensor in buffer solution 2.
14. When measuring without temperature sensor:
Enter the temperature of the buffer with
<
><
>
.
15. Start the measurement with
<ENTER>
.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter
flashes.
16. Wait for the measurement with stability control to be completed or ter-
minate the stability control and take over the calibration value with
<ENTER>
.
The pH value of the buffer solution is displayed.
17. Set the nominal buffer value for the measured temperature with
<
><
>
.
18. Accept the calibration value with
<ENTER>
.
The calibration display for the next buffer appears (voltage display).
19. If necessary, finish the calibration procedure as a two-point calibration
with
<M>
.
The calibration record is displayed.
01.08.2013 08:00