Operation
Multi 3410
59
ba75863y18
01/2015
4.7
ORP voltage
4.7.1
General information
You can measure the following parameters:
ORP [mV]
Note
The sensor connection and the USB-B (device) interface are galvani-
cally isolated. This enables interference-free measurements in
grounded measuring media.
Preparatory activities
Perform the following preparatory activities when you want to measure:
4.7.2
Measuring the ORP
Note
IDS ORP sensors are not calibrated. However, you can check IDS
ORP sensors using a test solution.
1
Connect the IDS ORP sensor to the meter.
The ORP measuring window is displayed.
2
Check the meter with the IDS ORP sensor.
1
Perform the preparatory activities according to section 4.6.1.
2
Immerse the IDS ORP sensor in the test sample.
3
Select the mV display with
<M>
.
15.03.2014 08:00