DIQ/S 281
Outputs
109
ba77244d01
08/2020
The control parameters result in the following (negative) characteristic curve:
Imin
8 mA
Imax
14 mA
Io
12 mA
ti
0 s (no I algorithm)
td
0 s (no D algorithm)
figure 6-18 Characteristic curve for application example
0
8
14
12
Imax
Imin
Current I [mA]
Measured value [mg/l]
4
2
8
Io
Nominal value
Xp