87
3.3.4
Quick Access to the devices via TitriSoft
The quick access menu (Fig. 83) can be used to send important direct commands to the TO 7280, TW 7650 and
the TitroLine
®
7500 KF
trace
.
Fig. 83
1) TitroLine
®
7500 KF
trace
Fig. 84
You have the following options:
«
Pump on
»:
Manually switches on the pump when using air as carrier gas.
«
Pump off
»:
Manually switches off the pump when using air as carrier gas.
«
Gas on
»:
Manually switches on the valve, the nitrogen is supplied to the cell.
«
Gas off
»:
Turns the valve off manually, the nitrogen is no longer supplied to the cell.
2) TO 7280
Fig. 85
You have the following options:
«
Init
»:
Starts a re-initialization of the TO 7280.
«
Head Up
»:
Moves the needle head to the top position. The gas supply must be switched separately.
«
Head Down
»: Moves the needle head to the lowest position for sampling.
The gas supply must beswitched separately.