pH value
HandyLab 600
36
ba77064e04
08/2021
sor evaluation.
Carrying out a QSC
control calibration
A QSC control calibration can, e.g. be useful if the sensor evaluation noticeably
changed (after some normal calibrations).
You can carry out QSC control calibrations at greater intervals than normal cali-
brations.
1 QSC scale
The double arrow on the QSC scale indicates the current sensor
evaluation
1
25.0
7.007
pH
°C
0
14
01.02.2014
08:00
Info
USB output
1.
Open the menu for measurement settings with
<ENTER>
.
2.
In the
QSC
menu, select
Control calibration
with
<
V
><
W
>
.
The calibration display appears.
AutoCal QSC-Kit
is displayed as the
buffer.
Exclusively use the QSC-Kit for the QSC calibration. If you use other
buffers, you will have no valid QSC control calibration.
3.
Follow the user guide.
The calibration is carried out like a normal three-point calibration. As
soon as the three-point calibration has been successfully carried out
you can decide whether to accept or discard the calibration as the QSC
control calibration.