HandyLab 600
pH value
ba77064e04
08/2021
25
Continuing with two-
point calibration
4.
Thoroughly rinse the sensor with deionized water.
5.
Immerse the sensor in the first buffer solution.
6.
For measurements without temperature sensor :
Enter the temperature of the buffer with
<
V
><
W
>
.
7.
Start the measurement with
<ENTER>
.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter
flashes.
8.
Wait for the end of the measurement with stability control or accept the
calibration value with
<ENTER>
.
The calibration display for the next buffer appears (voltage display).
9.
If necessary, finish the calibration procedure as a single-point calibra-
tion with
<M>
.
The calibration record is displayed.
For
single-point calibration
, the instrument uses the Nernst slope
(-59.2 mV/pH at 25 °C) and determines the zero point of the IDS
pH sensor.
pH
1
-180.0
24.8 °C
01.07.2011
08:00
Buffer
mV
pH
-180.0
24.8 °C
01.07.2011
08:00
AR
Buffer
mV
10. Thoroughly rinse the sensor with deionized water.
11. Immerse the pH sensor in buffer solution 2.