HandyLab 100
Operation
35
ba77062d02
12/2017
is automatically finished with the measurement of buffer solution 1 and
the calibration record is displayed.
1
In the measured value display, select the measured parameter
pH or mV with
<M>
.
2
Start the calibration with
<CAL>
.
The calibration display for the first buffer appears (voltage dis-
play).
3
Thoroughly rinse the electrode with deionized water.
4
Immerse the electrode in buffer solution 1.
5
When measuring without temperature sensor:
Measure the temperature of the buffer manually and enter it
with
<
><
>
.
6
Start the measurement with
<ENTER>
.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parame-
ter flashes.
7
Wait for the end of the measurement with stability control or
accept the calibration value with
<ENTER>
.
pH
1
-180.0
24.8 °C
AutoCal TEC
15.03.2014
08:00
Buffer
mV
pH
-180.0
24.8 °C
AutoCal TEC
10.011
15.03.2014
08:00
AR
Buffer
mV