9
SW 838023
EN 2.10 Shuttle Manual 1712
Mounting
Electrical mounting
Level transmitter
The Shuttle
®
must not be connected to
the power supply before the ultrasonic
sensor is mounted and connected cor-
rectly.
When the cover has been removed, the
green plastic film with the menu symbols is
tipped up to gain access to the terminals.
Shuttle
®
can be supplied with 10 - 30 V DC
on terminal 10 and 12 or with 230 / 115 V
AC on terminal L and N.
Current regulations for conductor and
fuse dimensions should always be ob-
served.
Always confirm that the
Shuttle
®
voltage
rating match the present voltage.
If
Shuttle
®
is delivered for 115 V AC supply,
it will be indicated with a label below the
leftmost terminal block as shown here:
Shuttle
®
is intended
Shuttle
®
is intended
for
230 V AC
. for
115 V AC
.
Terminal: Designation:
1 - 5
Ultrasonic sensor
6 - 7
Relay output 1
(Max. 50 V, 1 A resistive load)
8 - 9
Relay output 2
(Max. 50 V, 1 A resistive load)
10 and 11 4-20 mA output
(Max. 500
Ω
load)
10 and 12 10 - 30 V DC supply
L
230 / 115 V AC live
N
230 / 115 V AC neutral
Mount the wires according to the terminal
numbers on the reverse side of the green
plastic film: