background image

 

User manual EBI 330 

V 1.1 

Page 6 / 16 

Unpacking / Package Contents 

 

After receipt of delivery, please check the equipment for 
any damage. Furthermore, please verify that the delivery 
corresponds to your order. 

 

Scope of Delivery 

The delivery includes: 

x

  10 EBI 330 temperature data loggers, each with 

battery, protective cap and a label, on which the 
device type, serial number, bar code and 
temperature limits are printed.

 

Should you have a complaint for any reason, please do 
not hesitate to contact us: 

 

ebro Electronic GmbH 

 

Peringerstraße 10  D-85055 Ingolstadt 

Tel. +49 (0) 841 

 9 54 78 

 0 

Fax. +49 (0) 841 

 9 54 78 

 80 

Internet: www.ebro.com 

eMail: 

[email protected]

 

 

 

Summary of Contents for ebro EBI 300

Page 1: ...Single Use USB Data Logger EBI 330 T30 EBI 330 T85 User Manual...

Page 2: ...V 1 1 Page 2 16 Table of contents Overview 3 Safety instructions 5 Unpacking Package Contents 6 Configuration 7 Start Stop 8 LEDs 9 PDF 10 Battery 11 Cleaning and maintenance 11 Disposal 11 Technical...

Page 3: ...ecorded measurement values and the alarm status limit violations will be automatically generated The PDF file can be opened printed or archived without any other additional software The USB connection...

Page 4: ...User manual EBI 330 V 1 1 Page 4 16 1 2 3 1 Eye 2 LED 3 Start Stop button 4 USB connector 5 Protective cap 4 5...

Page 5: ...eas Danger of death Do not use the device in an environment hotter than 60 C The lithium battery may explode Do not expose the unit to microwave radiation The lithium battery may explode To ensure the...

Page 6: ...The delivery includes x 10 EBI 330 temperature data loggers each with battery protective cap and a label on which the device type serial number bar code and temperature limits are printed Should you h...

Page 7: ...e x Language of the PDF file You can apply these settings by downloading the free of charge software Winlog basic from http www ebi330 com and using the software to program the device Please refer to...

Page 8: ...ure and store temperature data according to the set interval until the data memory is full up to 6000 values the battery empty lasts for at least 100 days or the device is being stopped To stop the lo...

Page 9: ...reen LED will blink every five seconds This will continue until the logger is being stopped the battery is empty or a temperature limit has been violated Should any of temperature limits be violated f...

Page 10: ...ar to the ones of USB sticks The file can easily be copied from there The file has a lot of information to the measurement the logger configuration statistic data to and a chart of the measurement and...

Page 11: ...battery of the EBI 330 T85 loggers lasts for at least 100 days In neither logger version you can replace the battery Cleaning and maintenance Please use a damp cloth to clean the logger Do not use a s...

Page 12: ...DIN EN 12830 S T C D 1 ATP Calibration batch calibration Size 80 x 28 x 12 mm Weight 16 g incl battery EBI 330 T30 Temperature 30 C 60 C 22 F 140 F Accuracy 0 5 C 20 C 40 C 0 8 C for the remaining ran...

Page 13: ...e 13 16 EBI 330 T85 Temperature 85 C 50 C 121 F 122 F Accuracy 2 0 C 85 C 30 C 0 8 C 30 C 20 C 0 5 C 20 C 50 C Operating temp 85 C 50 C 121 F 122 F Storage temperature 40 C 85 C 40 F 185 F Power suppl...

Page 14: ...roduct complies to CE guidelines 2004 108 EG The product EBI 330 complies to the guidelines in accordance with EN 12830 Suitability S Storage T Transport Location C Food storage and distribution syste...

Page 15: ...User manual EBI 330 V 1 1 Page 15 16...

Reviews: