XTCERA X-MILL 500 User Manual Download Page 1

User Manual

X-MILL 500 Five-Axis Dental Milling machine

X-MILL 500

500

Address: 4th Floor,1st Building,2nd Area,Honghualing Industrial
               Park,Nanshan District, 518055, Shenzhen,China
Tel: 0086 755 8600 1804             Fax: 0086 755 8600 1486
E-mail: [email protected]        Web: en.xtcera.com

Shenzhen XTcera Medical Technology Co.,Ltd

Summary of Contents for X-MILL 500

Page 1: ...X MILL 500 500 Address 4th Floor 1st Building 2nd Area Honghualing Industrial Park Nanshan District 518055 Shenzhen China Tel 0086 755 8600 1804 Fax 0086 755 8600 1486 E mail xtcera xianton com Web e...

Page 2: ...fety of operators and the machine it should be operated by professional operators If you have any questions please contact us in a timely manner and our professionals will always be ready to help you...

Page 3: ...ur and has eight burs The maximum speed of spindle can be up to 60000 rpm which is extremely strong accurate and reliable The machine has strong adaptability and selects different processing strategie...

Page 4: ...llation Considerations 1 Open the packing box and check that the appearance of the equipment is damaged External dimension 720x850x860 mm Weight 225 kg Power supply voltage Single phase AC 220V 50 Hz...

Page 5: ...loor 2 2 EQUIPMENT COMMISSIONING 1 Turn on the equipment power switch The lighting lamp of the processing area and the screen will turn on 2 Run the XMILL220 and click MANUAL button to enter the manua...

Page 6: ...d tight the protection bar and click COLLET Then take out the protection bar and put T1 the biggest bur into it Check if the NO behind T on the screen is 1 or not like picture below If the NO is 1 the...

Page 7: ...point If you click stop button during the processing the system will not remember the current point and the system will start from the beginning point when you resume milling again 6 You can repeat th...

Page 8: ...e reference position 3 Use left hand to hold tight the milling bur before loosen the spindle chuck to avoid any injury from the ejected bur same as picture below Please remember the equip ment shall n...

Page 9: ...sed by bumping 3 No residual dust is allowed on the edge of the material when it is installed Be sure to clean it in order to avoid the breakage caused by uneven force during the installation 4 When t...

Page 10: ...s with clean tissue in order to avoid any exception alarm caused by dust buildup 4 2 SPECIAL NOTES 1 If you use our equipment to process products other than the above mentioned mate rials please make...

Page 11: ...nse or even cause system crash 7 Do not move the equipment after it is installed If it is necessary to move it please contact us You can only apply force to the bottom plate during the movement No for...

Page 12: ...adjusted and vacuum can be manually started and shut down CA Custom Wax high efficiency processing mode where the rate cannot be adjusted and the vacuum is started WA Wax Class A Wax conventional pro...