MVN Analyze as Master and as Noraxon TeleMyo slave
4.1
Start and Stop a recording
4.1.1
Hardware connections
•
In addition to the normal MTw hardware setup, connect the BNC connector from
the TeleMyo DTS to Sync OUT 1 on the Awinda Station.
•
Connect the USB connection between the TeleMyo DTS and the PC.
•
When switched on, the DTS will display “USB ready”.
4.1.2
Software Setup Noraxon MyoResearch
The example given below is for is gait analysis, measuring the medial gastrocnemius,
tibialis anterior, semitendinosus and the rectus femoris. For an 8-Channel EMG
system, Channel 9 is selected as the synchronisation line in the Noraxon MyoResearch
Software. If a 16 channel system is in use, this is the 17th channel.
Return to the main menu, under >Measuring Options, go to >Recording Options,
then>Triggering tab.
Figure 6 - Noraxon "Recording Options" screenshot.
Check the check box beside “Start Recording”;
•
Go to the
drop down menu beside “When Channel”, select “Sync”;
•
Select Rises Above (ensure that this is also the direction indicated on
the mini-receiver);
•
Input e.g. 0.5V
•
Longer than 5ms