XS Instruments OXY 7 Vio Instruction Manual Download Page 21

 

 

21   

 

 

 

Performing the measurement  

 

Remove the protective cap of electrode, rinse it with distilled water; dab it with paper towel and dip in the 

solution to analyse. Gently stir and wait until the value stability, when the icon  

 appears on the display, 

take the reading.  

 

Note

: the polarographic sensor tends to consume the oxygen, and thus reducing gradually the detected 

value.  Make sure that there is a minimum flow in the sample to analyse; if you are working in a laboratory, 

keep the sample stirred. 

 

Barometric pressure compensation 

 

Since the measurement of the partial pressure of the dissolved oxygen is related to the barometric pressure, 

this instrument is able to compensate each variation, thanks to the integrated barometric sensor.  

In order to view the barometric pressure measured by the instrument, press the key 

  during 

measurement and scroll through the measurement parameters: %O

2

 

 mg/l 

 mbar. 

 

 

Probe DO 7  maintenance 

If the instrument does not calibrate or the reading does not stabilize, it is necessary  to perform a 

maintenance of the probe. In order to perform the maintenance, follow these steps in order: 

 

Replacement of electrolyte 

 

Cleaning of anode and cathode 

 

Replacement of membrane 

 

 

Electrolyte replacement  

 

 

 

Remove the membrane cap from the sensor, check that it is not punctured or damaged; if it is 

intact, it can be used again, otherwise it must be replaced. 

 

 

Wash the membrane and the sensitive part of the sensor with distilled water; remove any salt 

residue 

and dry with paper towel.  

Be very careful when handling the sensor and the membrane. Falls, shocks or crushing can damage 

the sensor and/or membrane.

 

 

 

Fill the membrane cap with distilled water at half level and screw it on the sensor (pay attention in 

the fixing phase, since the membrane does not have to be screwed strongly up to end run, because 

it can be damaged). Stir gently, unscrew the membrane and empty it completely; in this way, any 

traces of water or dust will be eliminated. 

 

Refill the membrane with new electrolyte; this time, fill it completely and screw the sensor making 

sure that no air bubbles appear inside.  A light leakage  of electrolyte during the screwing of 

membrane, ensures that no air bubble will be created. 

 

Wash the probe and leave it in distilled water for at least half an hour, in order to rehydrate the 

membrane.  

 

Perform the 

Calibration of the sensor.

 If it does not work, proceed with the 

Maintenance of the anode and 

the cathode.

 

Summary of Contents for OXY 7 Vio

Page 1: ...1 Rev 1 0 11 02 2020 EN OXY 7 Vio INSTRUCTIONS MANUAL O2 mg l mbar Temp...

Page 2: ...2 Rev 1 0 11 02 2020 EN...

Page 3: ...tion 9 Display 9 Keyboard 10 LED 10 Installation 10 Supplied components 10 Start up 10 Turning on and off 11 Replacement of batteries 11 Instrument transportation 11 Key functions 11 Inputs Outputs co...

Page 4: ...oxygen measurement 20 Before starting 20 Measure Mode 20 Performing the measurement 21 Barometric pressure compensation 21 Probe DO 7 maintenance 21 Electrolyte replacement 21 Maintenance of the anode...

Page 5: ...this instrument is also suitable for use in the laboratory The innovative high definition colour LCD display shows all the necessary information such as the measurement the temperature the buffers use...

Page 6: ...zardous situation which could lead to serious injury or death if not avoided ATTENTION for a dangerous situation with reduced risk which can cause material damage data loss or minor or medium sized ac...

Page 7: ...ular functionality of the device and the operator safety are guaranteed only if all the following indications are respected the instrument can be used in accordance with the specifications mentioned a...

Page 8: ...e person who owns and uses the tool or authorizes its use by other people is the owner of the tool and is responsible for the safety of all users of the tool and third parties The owner of the instrum...

Page 9: ...Environmental operating conditions 0 60 C Maximum permissible humidity 95 non condensing Maximum altitude of use 2000 m System dimensions 185 x 85 x 45 mm System weight 400 g Reference regulations EMC...

Page 10: ...ument is always supplied inside the specific carrying case with these accessories batteries standard 0 oxygen paper tissues screwdriver beacker multilingual user manual and test report Contact your lo...

Page 11: ...the left compartment and two in the right compartment and insert the new ones Pay attention to the correct polarity Follow the diagram above the battery symbol in the rear compartment of the instrumen...

Page 12: ...with IMPORTANT When the Sleep mode is active by default after two minutes of inactivity of the instrument press any key to reactivate the brightness of the display Only at this point do the keys regai...

Page 13: ...before turning off To scroll through the different parameter screens press the key Sequence of parameters in measure mode OXY 7 Vio O2 mg l mbar Note Pressing the button after the last parameter the...

Page 14: ...ional keys and confirming with key OXY 7 Vio DO SETTINGS SETTINGS Within the selected menu move between the different programs using the directional buttons and press the button to access the submenu...

Page 15: ...til the value starts to flash then adjust it by continuing to use the directional keys press to confirm Note with the polarographic sensor supplied together with the instrument the manual temperature...

Page 16: ...ify the measurement by setting the indicative salinity value of the sample The default value is 0 ppt access the parameter SALT COMPENSATION P5 2 of the setup menu to change it and select the desired...

Page 17: ...sor uses an RCA Chinch connector Membrane cap Sensing element The permeable membrane allows the passage of only gas present in the sample to be analysed blocking the passage of liquids The oxygen reac...

Page 18: ...crew the membrane without electrolyte protecting the sensor with its rubber cap Oxygen sensor calibration The polarographic sensor is an active sensor which changes its response with wear and aging th...

Page 19: ...TTINGS P5 0 and confirm pressing the button again Press the button again confirm the entry into submenu CAL 0 P5 1 On the display the string POINT oxy 0 0 appears the device will search for the value...

Page 20: ...alibration line is out of the acceptable range 80 120 CALIBRATION TOO LONG The calibration exceeded the time limit only the points calibrated up to that moment will be kept Dissolved oxygen measuremen...

Page 21: ...Replacement of electrolyte Cleaning of anode and cathode Replacement of membrane Electrolyte replacement Remove the membrane cap from the sensor check that it is not punctured or damaged if it is inta...

Page 22: ...olyte If even after these procedures the probe does not calibrate replace the probe Instrument Setup Menu In measure mode press key to access the SETUP menu Use the directional keys to move to SETTING...

Page 23: ...o a minimum significantly saving battery consumption Note Sleep mode only affects the brightness of the display All other instrumental functions continue to operate normally To exit from the Sleep mod...

Page 24: ...ll repair or replace defective components This warranty does not apply if the product has been damaged used incorrectly exposed to radiation or corrosive substances if foreign materials have penetrate...

Reviews: