Page 1: ...User Manual BL 303 BL 602 BL 2002 BL 3001 BL 6001 BL 30K1 Models ...

Page 2: ...6 START 7 DATE AND TIME SETTINGS 7 USER CALIBRATION 7 CALIBRATION PROCEDURE 7 UNIT SETTING 8 UNIT SELECTION 8 DEFAULT UNIT SETTINGS 8 ZERO 8 TARE 8 MEASURE 8 COUNTER FUNCTION 9 RS 232 SETTINGS 9 ITEM SETTINGS 9 BATTERY 10 TROUBLESHOOTING 10 DATASHEETS 10 DISPOSAL OF ELECTRONIC DEVICES 11 ...

Page 3: ...3 Glass protection BL 303 ...

Page 4: ...4 BL 602 1 2 3 4 5 6 7 8 ...

Page 5: ...5 BL 2002 BL 3001 BL 6001 ...

Page 6: ...C Working Humidity 40 85 Installation and removal of the scale must be done with the machine switched off Do not use the scale upside down After switch On wait at least 30 minutes warming up before the first weighing For assistance contact your dealer or an authorized service center Keyboard On Off Hold for 3 seconds to start Calibration procedure Unit selection Print data RS 232 printer PC Tare Z...

Page 7: ...he value scroll to the next value confirm Successively are required 1 Month Month 2 Day Day 3 Hour Hour 4 Minute Min User Calibration Each balance is calibrated in the laboratory by technicians using certified sample masses Provided with all models there is a sample mass useful to test the functionality of the instrument Calibration procedure o Turn on the balance and wait at least 30 min of warmu...

Page 8: ...With no weight on the scales press the button to take the 0 00 reading on the display The ZERO function is only effective if the value shown is less than 4 of total capacity If the value is greater than 4 the load is regarded as a tare Tare After placing the tare on the scales press the button to reset the value displayed to 0 00 The load used for Tara has to be subtracted by the overall capacity ...

Page 9: ...the display shows first rS Set 8888888 According with the following table set the correct values a 0 9600 1 19200 b 0 print in continuos 1 print when stable 2 press to print c 0 1 print the date d 0 1 print the item e 0 PC Mode 1 special paper 2 continuos paper 3 micro printer f 0 g 0 Confirm with button Item settings Is possible to set an Item that will be printed with the other values for exampl...

Page 10: ...rument in correspondence of the load cell Datasheets Models BL 303 BL 602 BL 2002 BL 3001 BL 6001 BL 30K1 Cod 27000513 27000323 27000313 27000113 27000123 27000133 Capacity 300 g 600 g 2000 g 3000 g 6000 g 30 000 g Division 0 001 g 0 01 g 0 01 g 0 1 g 0 1 g 0 1 g Repeatability 0 001 g 0 01 g 0 01 g 0 1 g 0 1 g 0 2 g Linearity 0 003 0 03 0 02 0 2 0 2 0 2 Sensor Strain gage Strain gage Strain gage S...

Page 11: ...d of in public landfills According to the UE Directive 2002 96 EC the European users of electrical and electronic equipment can return it to the dealer or manufacturer upon purchase of a new one The illegal disposal of electrical and electronic equipment is punished with an administrative fine Ver 1 0_ENG_21 04 2015 ...