background image

XO CARE A/S  Usserød Mølle Håndværkersvinget 6  DK-2970 Hørsholm Denmark

tel +45 70 20 55 11 fax +45 70 20 55 10 www.xo-care.com

XO ODONTOSURGE

Instruction Manual

Y

A-551, vers. 1.00

Summary of Contents for ODONTOSURGE

Page 1: ...XO CAREA S Usser d M lle H ndv rkersvinget 6 DK 2970 H rsholm Denmark tel 45 70 20 55 11 fax 45 70 20 55 10 www xo care com XO ODONTOSURGE Instruction Manual YA 551 vers 1 00...

Page 2: ...followed in order to enhance safety and effectiveness In particular for those with little or no experience with HF Soft Tissue Handling and also for those familiar with conventional low frequency uni...

Page 3: ...N OFF Finger Switch 6 E How to Operate 7 F Preparation and Familiarization 9 F1 Dissection Exercise 9 F2 Coagulation Exercise 9 F3 GeneralAdvice 9 G Clinical Application 11 G1 Contraindications to HF...

Page 4: ...OSURGE 1 Instrument socket 5 Handpiece holder 9 Mains power cord 2 ON OFF finger switch 6 Mode switch panel for instrument activation 3 Handpiece detachable 7 Seven instruments in holder casing 4 Hand...

Page 5: ...10 green lamp 11 blue lamp 12 yellow lamp COAG CUT 13 Handpiece cord 15 Mains power inlet 17 Fuse drawer connector 14 Warning symbol 16 Main ON OFF switch see chapter M 3...

Page 6: ...tral role for the completely automatic and instantly re sponding power tuning system of the XO ODONTOSURGE As a result the unique power tuning system is continuously controlling and precisely adapting...

Page 7: ...the handpiece holder by inserting its tongue into the slot in the side of the control box left or right side at option C3 Control Box 1 Check that your unit is correctly rated for your local power lin...

Page 8: ...ady to operate Blue lamp OFF means standby mode ST BY mode with handpiece inoperative CUT switch Turns yellow lamp on and off Yellow lamp ON means cutting mode CUT mode with handpiece ready to operate...

Page 9: ...ment is now ready for operation Activate the instrument by pressing and holding the ON OFF finger switch on the handpiece Note the acoustic signal for cutting Coagulation haemostasis 1 As a first step...

Page 10: ...N OFF finger switch responds with an acoustic signal Do not touch the instrument when the acoustic signal is sounding CAUTION Always switch to standby mode only green lamp on as soon as the treatment...

Page 11: ...not be dry If it is moisten lightly with water 5 Hold the handpiece in a pencil type grip and hold the instrument in position slightly above the tissue surface Activate the instrument by pressing and...

Page 12: ...essing and holding the ON OFF finger switch The acoustic signal for the COAG mode will now be heard 4 Make repeated taps with the activated instrument on the tissue This will result in small target ar...

Page 13: ...rative dentistry will greatly benefit from your use of the XO ODONTOSURGE Special instructions are available upon request G1 Contraindications to HF Soft Tissue Handling 1 HF Soft Tissue Handling is c...

Page 14: ...o autoclave the instruments between pa tients The instruments can be sterilized by steam autoclaving maximum temperature 135 C 271 F conveniently when placed in the holder casing with opened or remove...

Page 15: ...accordance with this Manual The XO ODONTOSURGE and its accessories must not be taken into use if signs of operational electrical or mechanical defects are found The XO ODONTOSURGE must not be taken in...

Page 16: ...M D D 93 42 EEC Primary fuses For mains voltage 100 and 115V 2 00 amp time lag For mains voltage 230V 0 80 amp Generator frequency 27 MHz Max output power 50 watts at non inductive load impedance of 5...

Page 17: ...s to the safety reliability and performance of the XO ODONTOSURGE will not be accepted by XO CARE A S a If the equipment has not been used under normal conditions for its intended purposes and in stri...

Page 18: ...s whether they are in the form of an anesthetic or skin preparation agent are produced by natural processes within body cavities or originate in surgical drapes tracheal tubes or other materials 3 The...

Page 19: ...o a serious deterioriation of the state of health of a patient or user In the improbable event that your XO ODONTOSURGE should be involved in such an inci dent it is therefore very important that you...

Reviews: