![background image](http://html1.mh-extra.com/html/xo/flex/flex_installation-instructions-manual_3545969011.webp)
XO FLEX – Installation instructions
Necessary
tools
11
3
NECESSARY TOOLS
The following tools are necessary for installing, adjusting and servicing the XO FLEX unit:
3.1
STANDARD TOOLS
•
Hammer drill for at least 18 mm concrete drill bit (SDS-system recommended)
•
Carbide drill bit for concrete 18 mm (SDS-system)
•
Carbide drill bit for concrete 10 mm (SDS-system)
•
Wood bit 6 mm and 8 mm
•
Spirit level 20 – 30 cm
•
Combination wrench 17, 19 and 22 mm
•
Socket spanner with extension 19 mm
•
Torx screwdriver: T8, T9, T10, T15, T20, T25, T30 and T40
•
Pozidrive screwdriver: PZ 1-2-3
•
Allen key set: 1.5 – 10 mm
•
Digital multimeter
3.2
SPECIAL TOOLS SUPPLIED WITH THE UNIT
•
Special wrench (MG-416)
3.3
NECESSARY SPECIAL TOOLS AVAILABLE FROM XO CARE
•
Manometer, four holes, Midwest – for adjustment and control of drive air pressure to air
instrument (turbine) (FA-041)
•
Four carrying handles for the unit stand (AN-181)
•
Flowmeter – for adjustment and control of cooling air for motor (FA-400)
•
Antistatic wristband (UC-600)
•
Drill template 1:1 (AP-919)
3.4
OTHER SPECIAL NECESSARY TOOLS
•
Torque wrench 75 Nm
•
Two component flame retardant foam: Würth part nr 0893 303 200