If much blood has been suctioned through the sy-
stem, we recommend that the system be flushed with at
least 1 litre of cold water. After this procedure, the system
can be disinfected with XO Suction Disinfection.
xxx
Figure 16 Removal of suction nipple covers
xxx
Figure 17 Flushing of suction system
1 Container with XO Suction Disinfection
2 Push button, start Suction Disinfection
3 Suction nipples
The display shows the progress of the disinfection process.
When “SUCTION OK” appears in the display, the process has
been carried out successfully.
Please note that manual cancellation of flushing is
not possible.
XO Suction Disinfection container (1) must be replaced at spe-
cific intervals. When the contents in the container are under a
certain level, a “warning” will be activated and the message
“CHECK YELLOW BOTTLE” appears in the unit display.
XO Suction Disinfection is a corrosive liquid. Read
the safety data sheet supplied with the product.
Be careful not to spill the fluid. Any spills must be
wiped up at once. Then clean the surface with a damp cloth.