Xo 4 User Manual Download Page 15

XO

® 

4 User Guide

YB-253 Version 3.10

8

Figure 15  XO Water Clean cartridge

4.2.2

Single step water treatment (heavy de-
posits)

The “single step” water treatment should be used when the bac-
terial load of the unit water lines is considered to be too high, for 
instance, after vacation or other longer periods without usage.

We recommend that the single step water treatment be perfor-
med 3 consecutive times after such periods.

1 Place all water-bearing unit instruments in the instrument 

holder on the cuspidor. Place the tube with the “T-piece” on 
the cup filler outlet.

2 Activate 

 (not 

). The expansion container is emp-

tied and the water lines are flushed.

The display will show the remaining time of the disinfection pro-
cess. When "FLUSH OK" appears in the display, the process 
has been carried out successfully.

 Please note that manual cancellation of the flushing 

and expansion container water discharge process is not 
possible.

4.3

Disinfecting the suction system

The suction valves and tubes inside the unit are constantly flus-
hed with clean water while suction is activated. However, in or-
der to maintain an aseptic working environment, the suction 
system needs to be cleaned and disinfected daily.

Disinfection and cleaning of the suction system is also important 
to ensure problem-free operation of the equipment. Otherwise, 
the suction system may clog up, and break down as a result.

If you have chosen to equip XO 4 with XO Suction Disinfection, 
please proceed to section 4.3.1.

If your XO 4 does not have XO Suction Disinfection, the system 
must be cleaned manually. 

We recommend the use of the Dürr OroCup system with Dürr 
Orotol Plus suction rinse. 

If the suction tubes or drain contains heavy deposits, we recom-
mend the use of Dürr MD 555 Orotol® special cleaner combined 
with the Dürr OroCup.

In either case, please be sure to follow the instructions given by 
Dürr.

 Do not use any other disinfectants or cleaning pro-

ducts! Please note that failure to comply with this will affect 
XO’s warranty liability.

4.3.1

Flushing with XO Suction Disinfection 
(optional)

Every day the suction system (valves and tubes) must be disin-
fected by flushing with XO Suction Disinfection (ref.no. AN-354).

• Disconnect the suction tubes and press the button (2) (Figu-

re 17) once to deactivate suction.

• Remove covers from suction nipples.
• Place the two suction nipples on the disinfection nozzles (3) 

and reactivate the push button (2), following which flushing 
starts.

The process takes approximately 6 minutes, depending on the 
suction power.

The suction nipple covers can be autoclaved at max. 134 

o

C.

1

Summary of Contents for 4

Page 1: ...XO CARE A S H ndv rkersvinget 6 DK 2970 H rsholm Denmark Tel 45 7020 5511 www xo care com Valid from Firmware version 3 00 0470 XO 4 Workstation User Guide YB 253 Version 3 10...

Page 2: ......

Page 3: ...acceptance has been given by XO CARE A S hereinafter called XO 1 3 Rights Instruments and accessories built into XO 4 or used in combina tion with XO 4 by third party medical device manufacturers are...

Page 4: ...ositioning 5 3 3 3 XO 4 Lamp 5 3 3 4 XO HD Display 6 3 3 5 Suction 6 3 3 6 Call assistant 6 4 Infection control 7 4 1 Disinfectants recommended for XO 4 7 4 2 Disinfecting unit water lines with XO Wat...

Page 5: ...separator 14 5 3 Instrument hoses 14 5 4 Bien Air MC3 micro motor 15 5 5 Ultrasonic scaler 16 5 6 Light curing lamp 16 5 7 XO 4 Lamp halogen type 16 5 8 XO 4 Foot Control 16 5 9 Hand instrument table...

Page 6: ...5 XO Water Clean cartridge 8 Figure 16 Removal of suction nipple covers 9 Figure 17 Flushing of suction system 9 Figure 18 Removal of suction hoses and filters 10 Figure 19 Press the button to eject f...

Page 7: ...sion 3 10 vii List of tables Table 1 Sounds 2 Table 2 Error messages 17 Table 3 Service messages 18 Table 4 Consumables 19 Table 5 Useful spareparts 19 Table 6 Useful infection control parts 19 Table...

Page 8: ...tions in order to carry out easy and quicker treatment with optimum patient comfort The infection control section describes how you maintain the highest level of infection control while treating the e...

Page 9: ...rument is activated using the XO Foot Control If more than one the first instrument is lifted forward the other instruments are blocked If a second instrument is lifted forward while the first instrum...

Page 10: ...ct heated non heated water xxx Figure 3 Multi functional syringe 3 2 4 Micro motor The micro motor is operated as described in XO 4 Quick Guide The speed adjusts exponentially in relation to pedal mov...

Page 11: ...ument brid ge is included in XO 4 s range of accessories This is particularly suitable for dentists working solo The hand instrument table is available in 2 sizes A standard size for 1 cassette and a...

Page 12: ...neck and head 5 When the backrest passes approx 45 degrees lock the headrest again see Figure 8 6 Now only minor adjustments need to be made to adjust the headrest optimally to the patient xxx Figure...

Page 13: ...relieved by pressing them into their slots When working solo move the suction hose holder forward to fa cilitate easy access to the suction hoses Figure 12 Suction model 2005 Working DUO Figure 13 Su...

Page 14: ...g with the T piece on the cup filler outlet 2 Activate the expansion container is now emptied 3 Wait for 3 minutes The display will show the remaining time of the process 4 Switch off the unit and lea...

Page 15: ...cted daily Disinfection and cleaning of the suction system is also important to ensure problem free operation of the equipment Otherwise the suction system may clog up and break down as a result If yo...

Page 16: ...ay shows the progress of the disinfection process When SUCTION OK appears in the display the process has been carried out successfully Please note that manual cancellation of flushing is not possible...

Page 17: ...sh the filters in a thermodisinfector 90 o C Lubricate the O rings on the filters with XO silicone grease ref no YR 002 before inserting clean filters ref no MR 075 If air starts to leak from the suct...

Page 18: ...ce is replaced 4 4 2 XO 4 Lamp The splash screen may be removed for cleaning by the two screws Clean and disinfect with soap Use a clean cloth The reflector can be cleaned carefully using tap water an...

Page 19: ...claves may in principle be applied under the following guidelines cf EN 13060 Type N gravitation autoclave Massive unpacked material Type B vacuum autoclave Non massive packed material and fabrics Sin...

Page 20: ...joint 2 xxx Figure 25 Removal of syringe cover Do not wash the multifunctional syringe in a thermo disinfector and do not autoclave 4 5 4 Bien Air MC3 micro motor The micro motor cover 1 can easily be...

Page 21: ...igure 27 Access to amalgam separator Every 6 9 months the amalgam container must be replaced A warning signal activates when the container is 95 full xxx Figure 28 D rr CAS 1 COMBI SEPARATOR When the...

Page 22: ...new tube xxx Figure 30 Check for identical numbers In the event that the instrument hoses have been removed one must ensure that the plugs and sockets are completely dry before the hoses are re mount...

Page 23: ...gloves or a clean piece of cloth when handling the bulb 5 Use the plastic tool to carefully mount the bulb in the socket 6 Mount the shade and the splash screen again 7 Switch on XO 4 and activate XO...

Page 24: ...layed after start up of unit if no communication with Power Driver is detected Call your XO service provider STAND CONTROL FAIL CALL SERVICE Displayed after start up of unit if no communication with S...

Page 25: ...shut off power or activate any buttons FLUSH OK Water line flushing completed Unit operational again NEXT SERVICE YYYY MM DD Less than 30 days to next recommended preventi ve service Call your service...

Page 26: ...MD 555 Orotol Removal of heavy deposits from unit suction system n a D rr FD 322 surface disinfection Surface disinfection n a D rr FD 360 skai disinfection Upholstery cleaning disinfection n a Table...

Page 27: ...onal conditions for XO 4 we re commend the use of an oil free compressor s equipped with a dry air aggregate XO 4 fulfils the requirements in EN 1717 concerning the back flow of water if it is configu...

Page 28: ...he configuration of the unit when supplied from the factory See XO 4 Configuration Guide XO Foot Con trol The XO Foot Control is placed on the floor and consists of a pe dal and a joystick as well as...

Page 29: ...XO 4 User Guide YB 253 Version 3 10 22...

Page 30: ...is a registered trademark Printing errors if any are subject to reservations and XO CARE A S reserves the right at any time to alter specifications without further notice 2014 All rights reserved XO C...

Reviews: