XY
series precision electronic balance is a kind of intellectual measuring instrument with
weighing, counting and other many functions.
Ⅰ
Model specification and index
Model
Capacity
Readability
Size of weighing Pan
XY-2C
100g-3100g
0.01g
Ф130mm
XY-1B/BF
100g-5100g
0.1g
Ф130mm/160x160mm
Ⅱ
Unit change
There are four unit can be choose ,press
”
UNIT
”
and choose the unit you
need:
g,ct,oz,gsm
Ⅲ
Minimum weight setting
1. Under the weighing range you need, you can choose the minimum weight as your
requirement ,for example ,we choose the “0.00” accurate ,so we can set the minimum as
following:
1.
Turn on the balance when it display”0.00”
2.
Press”0”+”COU”,when it will display “1d”(there are 1d to 5d five digits can choose)
3.
Press “COU” or ” CAL ” to choose the minimum weighing you required.
Ⅳ
Display Icon
1.
Status icon :”+”and “-“is the reading state ,the ”o” means the weighing
process ,when it disappear ,the weighing is stable.
2.
The digits :the digits means the object weight
3.
The unit icon
4.The residual capacity of the battery(option)
Ⅴ
Working Condition
Working Temperature: 0°
~
40°C
,
Maximum Consumed Power: 5W
Temperature Fluctuation: 5°C/h
,
Power Supply: 100-220V, 50HZ/60HZ
Comparative Humidity: 50
﹪~
85
﹪
Ⅵ
Operation Requirements
When it is operated, electronic balance shall be placed on stable working platform to avoid
the influences from mechanical vibration, direct sunshine and air current etc.
Ⅶ
Operation
Power-on:Press the “ON/OFF”key, and the display indicates in turn “8.8.8.8.8.8.” “Max
weighing capability of balance”, “=, =, =, =, =”. At last it shall display the weighing mode as
“XXX.XX”. If any error founded in display, you may turn on and turn off the balance once
more. Then it will be operated normally.
Ⅷ
Calibration
(1)
Calibration Requirements
When distinct errors appeared in the weighing of balance, the balance shall be recalibrated
to make weighing accurate.The recalibrated balance shall be placed on the stable working
platform without the influences from the vibration of air current and strong electromagnetic
wave.The calibration results will be more accurate 20 minutes later after the balance is
operated.
(2)
Calibration Procedure
One-point Calibration
Turn on the balance it shall display the weighing mode as “0.00”, then press the “Cal”
button until shows “CAL”, then release. Display blinkingly the stand weight value and then
place same value standard weight; The “= = = = = =” will be displayed; then display stably
weight value and remove weight. then “= = = = = =” will be displayed. the “0.00” will be
displayed;. The calibration is finished.
Three-point Calibration
Turn off the balance,then press the“Cal”button and turn on the balance simultaneously until
the display shows “CAL”,then release.The calibration procedure shall be undertaken as
blow:
Display blinkingly the first point standard weight value then place the same value weight;
Display stably weight value and then remove weight. The “= = = = = =” will be displayed;
Display blinkingly the second point weight value and then place the same value weight;
Display stably weight value and then remove weight. The “= = = = = =” will be displayed;