XY 

series precision electronic balance is a kind of intellectual measuring instrument with 

 

weighing, counting and other many functions. 

 

 Model specification and index

 

Model

 

Capacity

 

Readability

 

Size of weighing Pan

 

XY-2C

 

100g-3100g

 

0.01g

 

Ф130mm

 

XY-1B/BF

 

100g-5100g

 

0.1g

 

Ф130mm/160x160mm

 

 Unit change

 

There  are  four  unit  can  be  choose  ,press 

UNIT

  and  choose  the  unit  you 

need:

g,ct,oz,gsm

 

  Minimum weight setting

 

1.    Under  the  weighing  range  you  need,  you  can  choose  the  minimum  weight  as  your 

requirement ,for example ,we choose the “0.00” accurate ,so we can set the minimum as 

following:

 

1.

 

Turn on the balance when it display”0.00”

 

2.

 

Press”0”+”COU”,when it will display “1d”(there are 1d to 5d five digits can choose)

 

3.

 

Press “COU” or ” CAL ” to choose the minimum weighing you required.

 

  Display Icon

 

 

1.

 

Status  icon  :”+”and  “-“is  the  reading  state  ,the  ”o”  means  the  weighing 
process ,when it disappear ,the weighing is stable.

 

2.

 

The digits :the digits means the object weight

 

3.

 

The unit icon                  

4.The residual capacity of the battery(option)

 

  Working Condition

 

Working Temperature:   0°

40°C

      

Maximum Consumed Power: 5W 

 

Temperature Fluctuation:   5°C/h

       

Power Supply: 100-220V, 50HZ/60HZ

 

Comparative Humidity:     50

﹪~

85

 

  Operation Requirements

 

When it is operated, electronic balance shall be placed on stable working platform to avoid 

the influences from mechanical vibration, direct sunshine and air current etc.

 

  Operation

 

Power-on:Press  the  “ON/OFF”key,  and  the  display  indicates  in  turn  “8.8.8.8.8.8.”  “Max 

weighing capability of balance”, “=, =, =, =, =”. At last it shall display the weighing mode as 

“XXX.XX”. If any error founded in display, you may turn on and turn off the balance once 

more. Then it will be operated normally.

 

 Calibration

 

(1)

 

Calibration Requirements

 

When distinct errors appeared in the weighing of balance, the balance shall be recalibrated 

to make weighing accurate.The recalibrated balance shall be placed on the stable working 

platform without the influences from the vibration of air current and strong electromagnetic 

wave.The  calibration  results  will  be  more  accurate  20  minutes  later  after  the  balance  is 

operated. 

 

(2)

 

Calibration Procedure

 

One-point Calibration

 

Turn  on  the  balance  it  shall  display  the  weighing  mode  as  “0.00”,  then  press  the  “Cal” 

button until shows “CAL”, then release. Display blinkingly the stand weight value and then 

place same value standard weight; The “= = = = = =” will be displayed; then display stably 

weight value and remove weight.    then “= = = = = =” will be displayed. the “0.00” will be 

displayed;. The calibration is finished.

 

Three-point Calibration

 

Turn off the balance,then press the“Cal”button and turn on the balance simultaneously until 

the  display  shows  “CAL”,then  release.The  calibration  procedure  shall  be  undertaken  as 

blow:

 

Display blinkingly the first point standard weight value then place the same value weight;

 

Display stably weight value and then remove weight. The “= = = = = =” will be displayed;

 

Display blinkingly the second point weight value and then place the same value weight;

 

Display stably weight value and then remove weight. The “= = = = = =” will be displayed;

 