![Xenios DELTASTREAM HC Instructions For Use Manual Download Page 39](http://html1.mh-extra.com/html/xenios/deltastream-hc/deltastream-hc_instructions-for-use-manual_3255965039.webp)
39
13
QUICk gUIDE
WARNINg
The quick guide does not absolve the
user of the need to follow the full instruc-
tion manual.
The unit switches all functions off in the
case of high-priority alarms. Take it out
of service and hand it over to a service
technician to perform checks and restore
operating safety.
CAUTION
There is a risk that the patient’s
blood could be heated or cooled
excessively.
Monitor the patient’s body temperature
if the unit is used on the patient with a
oxygenator.
WARNINg Do not operate the unit if:
• The display has failed;
• Individual segments of the tempera-
ture display have failed (temperature
can no longer be read clearly);
• The red fault indicator (4) stays on
or does not come on at all (function
test);
• The beep continues to sound or does
not sound at all (function test);
• The unit does not respond when
buttons are pressed;
• When the unit is switched on or
during the function test, the unit
does not react as described in the in-
struction manual (see “6.2.2 Function
test” on page 19).