CLEANINg AND DISINFECTION
Ventilation slots
Check the ventilation slots on the sides of the unit for dirt
regularly, but at least every six months. Do not open the
device yourself.
Dust deposits in the unit will reduce the efficiency of the
system. Have any inside dirt removed by an authorized
service technician.
Company
Product name
Actve substance
Isopropanol
Isopropanol (70%)
Bode Chemie GmbH
Kohrsolin
TM
Sterilium
TM
Bacilol Plus
TM
Aldehyde
Propanol
Propanol
Ecolab Healthcare
GmbH
Incidin Schaum
TM
Incidin Spray Liquid
TM
Propanol/Ethanol
Propanol/Ethanol
B.Braun Melsungen AG
Meliseptol
TM
Propanol
Clorox Company USA
Clorox Healthcare
Bleach Germicidal
Wipes
TM
Hypochlorid
Schülke&Mayr
Mikrozid Liquid AF
Propanol/Ethanol
Metrex
Cavi Wipes
Isopropanol/Butox
yethanol
PDI
Super Sani Cloth
Isopropanol
Maybe some of the listed products are not available in
your country. Take care about the availability!
7.1.2
CHANgINg WATER AND DISINFECTINg THE WATER CIRCUIT
We recommend changing the water with the subsequent
disinfection of the water cycle as follows in order to kill
water-borne pathogens (including non-tuberculous my-
cobacteria) and other germs:
• After each therapy,
• As part of the initial commissioning,
• at latest 30 days after the start of therapy,
• in case of storage and / or prolonged non-use
• generally after 30 days, also with stored devices
• every 12 months as part of the safety inspection.
• Replace hose set and couplings every 12 months.
Filling the device during initial operation and recommissioning:
• Fill approximately 1 liter of sterile water, e.g. with
the help of a disinfected funnel into the water tank
of the device.
• Make sure that no foreign objects enter the water
tank.
• Observe the water level indicator during filling.
The water level should be slightly below the MAX
mark after filling.
• After filling, screw the screw cap of the water filler
neck back in until the screw cap seals.
• Check the microbial contamination regularly.
We recommend to test the water quality once a
month, nevertheless the system is in use or stored.
Emptying the device:
• Disconnect the device from the mains.
• Place a container (bucket, bowl, etc.) under the device
or place it on a drain (e.g. sink).
• Unscrew the cap of the water filler neck in sterile
condition.
• Drain the tank contents into the collector via the
connected hoses. To do this, remove the adapter
between the Hansen couplings. Make sure that the
hoses are also completely emptied afterwards.