3

 

Maintenance 

3.1

 

Cleaning  

RadTarge II can be cleaned by wiping it with a damp cloth. Do not immerse RadTarge II in liquid as 
the device is not completely waterproof. 

3.2

 

Calibration 

RadTarge II does not require user calibration to work properly. However, depending on the rules and 
regulations of your industry, you may be required to regularly calibrate your equipment for quality 
assurance purposes. Calibration may only be performed by X-Z LAB or an accredited calibration 
laboratory. 

3.3

 

Troubleshooting 

Problem 

Solutions 

Unable to turn on 

Check whether the power button was pressed for 
longer than two seconds 
Ensure the battery has been fully charged 

Automatic shutdown 

Ensure the battery has been fully charged 

Insufficient power indicator on 

Ensure the battery has completed charging 

Charging indicator light not blinking 

Ensure the MicroUSB cable is connected properly 
Plug charging cable to a different USB port 

Continuous alarm ringing 

Confirm that a radiation source is nearby 
Check alarm threshold settings 

Alarm is not ringing near radiation source 

Check alarm threshold settings 

If the problems listed above cannot be solved, please contact X-Z LAB. 

3.4

 

Warranty 

RadTarge II comes with a one (1) year limited warranty for parts and labor. Cosmetic defects are not 
eligible for repair. The device is not eligible for warranty if it has been subjected to improper use or 
intentional damage. 

While the device is still under warranty, users may purchase an additional year of coverage for 5% of 

the product’s price. The warranty can be extended up to two (2) years

, totaling to three (3) years of 

coverage past purchase date. 

Devices past warranty may be sent in to X-Z LAB for repair. Rates for parts and labor will be 
determined at the time of repair. 

 

 

Summary of Contents for RadTarge II D700

Page 1: ...RadTarge II D700 Electronic Personal Dosimeter Operator Manual Version 1 1 January 2017...

Page 2: ...X Z LAB Inc Subsidiary of Raycan Technology Co Ltd 2440 Camino Ramon Suite 264 San Ramon CA 94583 Phone 925 359 6908 Email contact x zlab com www x zlab com...

Page 3: ...g 9 2 2 Interfaces 9 2 2 1 Dose Rate Interface 9 2 2 2 Cumulative Dose Interface 9 2 3 Settings 10 2 3 1 Alarm Options 10 2 3 2 Threshold Settings 10 2 3 3 Dose Clearing 10 2 3 4 Date and Time Setting...

Page 4: ...alert the wearer of alarm conditions Intuitive navigation allows users to adjust settings in the field This pager like direct reading EPD accurately detects and measures radiation exposure for workers...

Page 5: ...3 1 2 Product Diagram MicroUSB port Clip Charging indicator Detector location Display Power button Alarm indicator Settings button...

Page 6: ...ble 80 dB at 12 in 30 cm Visual LED and display Vibrating Alarm Response Time 2 s Overload Display Activation when 100 mrem h 1 mSv h Overload indication up to 1 000 rem h 10 Sv h Electrical and Mecha...

Page 7: ...Tech Code Energy keV Energy Response LK35 30 82 00 LK55 48 89 01 LK70 60 112 74 LK100 87 107 03 LK125 109 83 78 LK170 148 90 12 137 Cs 662 100 00 60 Co 1250 84 82 0 20 40 60 80 100 120 0 200 400 600 8...

Page 8: ...quivalent Rate Gy h Relative Intrinsic Error 137 Cs 662 keV 60 5 4 62 121 0 0 93 242 0 1 03 484 0 1 00 968 0 0 29 0 00 5 00 10 00 15 00 20 00 25 00 30 00 0 200 400 600 800 1000 1200 keV Relative Intri...

Page 9: ...se XY Plane 137 Cs Direction Response 180 100 150 96 120 95 90 99 60 96 30 97 0 100 30 98 60 96 90 100 120 97 150 98 50 60 70 80 90 100 180 150 120 90 60 30 0 30 60 90 120 150 Angular Response XY Plan...

Page 10: ...Cs Direction Response 0 100 30 94 60 91 90 84 90 83 60 92 30 93 1 5 Package Contents Item Quantity RadTarge II D700 1 MicroUSB Cable 1 User Guide 1 50 60 70 80 90 100 0 30 60 90 90 60 30 Angular Resp...

Page 11: ...e Rate Interface The dose rate interface is the main interface after turning on the device it displays the dose equivalent rate Accurate to 0 001 mrem h when the reading is below 100 mrem h else to 0...

Page 12: ...ve onto the next alarm mode 2 3 2 Threshold Settings Enter setup mode and press the settings button three times AL will flash at the top of the display Press the power button to enter threshold settin...

Page 13: ...n interface 2 4 Data Transfer and Charging Turn on RadTarge II Connect it to your computer using the included MicroUSB cable Data from the device can be transferred via RadSuite Dose to manage persona...

Page 14: ...gin User enters username and password on the login screen Device login After connecting a device that has been assigned to a user the system automatically logs into the user s account Status Bar The s...

Page 15: ...s Edit user info Device management Add delete device Edit device parameters Data retrieval Check device data A default Super Admin account is provided for new user registration Username Administrator...

Page 16: ...ntains basic information about the device 2 5 4 Cumulative Dose Curve Click the sidebar toggle to show or hide the informational sidebar There are two data modes to choose from Hp 10 deep dose equival...

Page 17: ...ed Power on Device start time Power off Device shutoff time Alarm start Alarm start time Alarm threshold Alarm stop Alarm end time Clicking the date selection menu reveals a dropdown menu Users can ad...

Page 18: ...n on the top right to access the settings interface The function selection bar updates to settings icons The icons and their corresponding functions are shown below Icon Functions User Group Managemen...

Page 19: ...nt and device management Personal Information The personal information interface allows users to set their personal information including name gender job position email address and photo Button Functi...

Page 20: ...ete users on the user management interface Clicking add leads to the user creation interface Fill in the user s basic information and assign a user group Generally setting the user group to user is re...

Page 21: ...ed users icons and usernames Click on the user s photo to show the targeted user The device selection bar will show the selected user s device list To add a device under the selected user s account cl...

Page 22: ...te the process Time Settings The time settings interface is used for internal device real time clock RTC calibration After entering the interface the time selection menu displays the date and time the...

Page 23: ...dose equivalent equivalent dose at 10 mm below surface of body Dose equivalent rate threshold Hp 0 07 Dose equivalent threshold Shallow dose equivalent equivalent dose at 0 07 mm below surface of bod...

Page 24: ...mpleted charging Charging indicator light not blinking Ensure the MicroUSB cable is connected properly Plug charging cable to a different USB port Continuous alarm ringing Confirm that a radiation sou...

Page 25: ...23 Revision History Revision Date Description 0 1 September 22 2016 Document created 1 0 October 28 2016 Document released 1 1 January 12 2017 Operating temperature updated...

Reviews: