vipPAQ Hardware Installation Manual (Ed. 2c)
9 Completion
1. Mount the measuring heads
2. Adjust the position screws on the measurement head(s). With this the switch to
lock/unlock the head upon the tilting of the measurement head gets adjusted).
a. Loose slightly the screw beside the position screw (red arrow)
b. Rotate the position screw, until the slit (green marked) is vertical and
the metal bolt is on top.
c. Tighten the screw beside the position screw (red arrow)
3. Connect the appropriate cable to the measuring heads and controller unit.
4. Lay the cables to the measuring heads in the rail slot. Fasten the cables with the
enclosed O-rings approximately all 10 cm / 4 inches.
Upon this, please make sure the cable length at the head is enough long to move
the heads later on.
5. Connect the controller unit to the PC’s USB interface and to the power supply.
6. Tighten up the screws on the operating lever.
Maximal cable lengths:
•
Max. length PC – USB converter:
2 m
•
Max. length USB converter – Controller:
50 m
•
Max. length controller - Measurement head:
3 m
10