X-Rite DensiEye 100 and 700
DensiEye Reference Manual 4
43
9. Maintenance and Care
9.1 Reset
If the instrument is unresponsive, the microprocessor may be locked up. This can
happen after changing batteries or another disruption. Press the Reset button (red
button on the bottom side of the measuring head). The instrument restarts and
displays the the status of the DensiEye as follows:
•
In the main menu, the settings Auto Mode (DensiEye 700) or Density (DensiEye
100) are selected
•
All settings remain the same
Procedure:
•
Press the Reset button for approximately 3 seconds
9.2 Total
Reset
After a Total Reset the status of the DensiEye is as follows:
•
In the main menu, the settings Auto Mode (DensiEye 700) or Density (DensiEye
100) are selected
•
All settings are reset to factory defaults
•
The reference and tolerance values remain the same
•
The calibration is reset
•
The device must be recalibrated
Procedure:
•
Press the Reset button
•
Press the Enter button
•
Release the Reset button
•
Release the Enter button