Measuring
NOTE:
Before procedure with a skin measurement, make sure to sanitize your hands and
the measurement surface of the instrument. Refer to the Sanitizing the Instrument in the
Appendices for the procedure.
1.
Make sure the white reference slider is in the storage location and position the CAPSURE
Cosmetic on the skin as previously explained.
2.
Lightly press and hold the
Measure
button to activate the preview. The area to be
measured is indicated in the preview by the circle.
3.
When you have located the area you wish to measure, press the
Measure
button firmly. An
audible “camera-click” sound indicates the start of the measurement. A progress bar
appears at the bottom of the display to indicate the measurement status. It is important
that the instrument is held steady for the duration of the measurement.
NOTE:
If averaging is activated in the Settings menu, a simulated color patch for the
measurement appears. Continue with Step 3 for averaging information, or advance to Step
4 if averaging is not used.
This screen indicates that two measurements are required. If you want the end averaging,
Press the
Enter
(Select) button.
Continue with additional measurements to complete the averaging procedure.
4.
After the measurement(s) is complete, the instrument displays the Filter screen or match
results. Match results initially appears if only one match is found. If more than one match is
found to be acceptable, the screen changes to the Filter screen. If no match is found to be
acceptable, “No close match found” message appears. Continue with Step 5 for multiple
match (Filter) information, or advance to Step 8 for single match information.