Troubleshooting
Prior to contacting X-Rite support department for instrument problems, try the applicable solution(s) described
below. If the condition persists, contact us using one of the methods listed in the Service Information section.
Problem
Cause
Solution
Instrument not
responding.
Instrument is in power down mode.
Press the power button.
Battery pack is very low or bad.
Charge the battery.
If battery pack is faulty, replace using the
procedure in the Appendix.
No battery pack installed.
Install batteries or plug in AC adapter.
Reset the instrument. See Setting Mode.
Measurement error or
results appear
inaccurate.
Material being measured is
damaged (e.g. scratched)
Obtain new material.
Instrument requires calibration.
Refer to Calibration mode for procedure.
Calibration procedure
fails.
Calibration reference is dirty or
damaged.
Clean the reference per procedure in Appendix,
or replace if damaged.
Instrument and software
not communicating (USB
connection).
Interface cable not connected.
Connect the interface cable between the
computer and the instrument.
Close and restart the software application. If this
does not work, reboot the computer.
Reset the instrument. See Settings Mode
procedure.
Instrument will not
measure or calibrate.
Sensor indicators in the display do
not change color.
Pressure sensors are not working correctly. Put
the instrument into targeting mode and place
the instrument on a flat surface and then lift. If
one or more of the sensor indicators in the
display do not change color when lifted, there
may be a problem with the pressure sensor.
Contact technical support.