G E T T I N G S T A R T E D
12
Calibrating the Instrument
Under normal circumstances, the instrument should be
calibrated at least once a day. The calibration procedure
consists of a white cal reading.
Make sure the calibration reference is clean before use.
Carefully clean the ceramic disk with a dry, lint-free cloth. Do
not use solvents or cleaners of any kind.
NOTE:
The instrument must be calibrated with the target window
removed when using the instrument with the shoe extended.
Calibration Procedure
1.
Press Tab Up
$
or Tab Down
@
to highlight
Calibrate
.
Press Enter key
#
to access the calibration mode.
2.
Position the target window on the white reference.
3.
Press the instrument firmly to the shoe. Hold steady until
the instrument indicates the white calibration is
completed.
4.
Store the calibration reference in a dry, dust free area,
away from direct exposure to light.