Troubleshooting
13
PROBLEM
CAUSE
REMEDY
Fluorescent bulb(s)
will not illuminate
Loose bulb(s)
Blown fuse
Loose connector
Faulty bulb(s)
Faulty switch
Faulty ballast(s)
Open wiring
Seat bulb(s)
Replace fuse
Re-seat connector
Replace bulb(s)
*
*
*
Reading will not
change or hold
Loose connector(s)
Faulty P.C. board
Faulty switch
Open wiring
Re-set connector
*
*
*
*
= Contact X-Rite Instrument Services or an authorized service center for
repair.