page 58 of 84
G S
PECIAL
F
UNCTIONS
G1 R
ESET TO
F
ACTORY PRE
-
SETTINGS
You can reset all adjustments and settings to the factory pre-settings. For this action press simultaneously the keys
<ENTER> and <ON/MODE> until an arrow to the right hand side appears in the display.
The following values will be set and the following actions performed:
BlueMETER:
•
Filter Type 3
•
Unit
mm/m
•
Display
mode
Absolute
•
Relative Base in millimetres, value 1000
•
Relative Base in Inch, value 10
•
All members of the wireless data transmission group are deleted
•
All members of the list of instruments are deleted
BlueLEVEL
•
Filter Type 3
•
Unit set to
o
1
μ
instrument:
mm/m, 3 decimals
o
5
μ
instrument:
mm/m, 3 decimals, rounded to 5µm/m
o
10
μ
instrument:
mm/m, 2 decimals
•
Display
mode
Absolute
•
Relative Base in millimetre, value 1000
•
Relative Base in Inch, value 10
•
All members of the wireless data transmission group are deleted
•
Relative Zero is set to 0
•
Absolute Zero is set to 0
G2 F
IRMWARE
V
ERSION
With a special key operation you can read the version number of the firmware installed. After turning off the
instrument hold the <ON/MODE> key down for another 10 seconds.
The display shows in large figures the date of issue and at the bottom the version number of the firmware.
Summary of Contents for BlueLEVEL
Page 2: ......
Page 68: ...page 68 of 84 L FLOWCHARTS BLUEMETER BASIC ONLY L1 POWER UP...
Page 69: ...page 69 of 84 L2 SAMPLING...
Page 70: ...page 70 of 84 L3 SENSOR...
Page 71: ...page 71 of 84 L4 REFRESH...
Page 72: ...page 72 of 84 L5 RELATIVE ZERO REL ZERO...
Page 73: ...page 73 of 84 L6 UNIT...
Page 74: ...page 74 of 84 L7 ABSOLUTE ZERO ABS ZERO...
Page 75: ...page 75 of 84 L8 FILTER...
Page 76: ...page 76 of 84 L9 RADIO TRANSMISSION...
Page 77: ...page 77 of 84 L10 JOIN...
Page 78: ...page 78 of 84 L11 LEAVE...
Page 79: ...page 79 of 84 L12 HOLD...