TEACH-IN of the IR-trigger (Zapper) to the BlueMETER BASIC and to the
BlueLEVEL
BASIC)
In order to eliminate interference of the zapper signals when several measuring groups are active in the triggering
range the IR trigger can be assigned to a specific measuring group by applying the function TEACH-IN
Procedure TEACH-IN:
•
The measuring or display instrument must be started
•
Keep the key <ON> on the measuring instrument (BlueLEVEL BASIC)
or
the key <ZERO/SELECT> on the display instrument (BlueMETER BASIC) pressed
•
Point the trigger (IR-zapper) in direction of the measuring or display instrument
•
Press
the
actuator key on the IR Zapper until both red IR LEDs are lighting up
When the instruments are dispatched this procedure is already factory set standard.