Technical data

TetraCon

®

700-...Ex

7

Technical data

7.1

General features

Measuring range

10 µS/cm ... 1000 mS/cm at 0 °C ... + 40 °C

Measuring principle

Four-electrodes measurement

Temperature sensor

Integrated NTC 30 (30 k

 / 25 °C)

Dimensions

Weight

Material

Connection cable

Applicable guidelines

and norms

Length

196 mm 
(without screwed cable gland)

Shaft diameter

40 mm

Approx. 660 g (without sensor connection cable)

Sensor head

PVC, epoxy (filler)

Electrodes, housing of the tem-
perature sensor 

Graphite

Shaft

POM, conductive

Closing head

POM, conductive

Protection ring

POM, conductive

Cable screw joint

POM

Cable coating

PUR

Length

Depending on variant:
– 1.5 m
– 7.0 m
– 15.0 m

Diameter

8.8 mm

Smallest allowed
bend radius

Permanent bend: 180 mm
Short time bend:

90 mm

Connection

Screw plug, 7 poles (IP 65)

General safety

– EN 60079-0
– EN 60079-11
– EN 61010-1

Summary of Contents for TetraCon 700-...Ex Series

Page 1: ...OPERATING MANUAL ba77055e01 05 2014 TetraCon 700 Ex Ex ib IIC T6 Gb X CONDUCTIVITY MEASURING CELL SUITABLE FOR OPERATION IN POTENTIALLY EXPLOSIVE ATMOSPHERES...

Page 2: ...caso di utilizzo in atmosfere potenzialmente esplosive osservare le disposizioni di sicurezza riportate nel documento ba77057y LT Lietuvi kai Dirbdami zonose su sprogia aplinka laikykit s ba77057y dok...

Page 3: ...7 2 2 3 Unauthorized use 7 3 Commissioning 8 3 1 Scope of delivery 8 3 2 Installation 8 3 2 1 Use in potentially explosive atmospheres 8 3 2 2 Connection to the measuring transmitter 8 3 3 Configurati...

Page 4: ...cture of the TetraCon 700 Ex conductivity measuring cell Fig 1 1 Structure of the TetraCon 700 Ex conductivity measuring cell 1 Shaft 2 Closing head with protective ring 3 Voltage electrodes 4 Tempera...

Page 5: ...influences from primary or secondary polarization effects This ensures a high degree of measuring accuracy The modern epoxy casting technique reduces the danger of breaking the conductivity measuring...

Page 6: ...indicates a possibly dangerous situation that causes death or serious injuries if the safety instruction is not followed WARNING WARNING indicates a possibly dangerous situation that can cause death o...

Page 7: ...operated according to the authorized use specified above The product may only be supplied with power by the energy sources mentioned in this operating manual The product may only be operated under the...

Page 8: ...sion Noncompliance with the safety requirements according to the explosion protection doc umentation ba77057y can cause a potentially explosive atmosphere to detonate Explosion protection is com plete...

Page 9: ...nsmitter The following values must be set in the CONF menu of the Stratos Pro measuring transmitter Menu item CONF menu Value Sensor type 4 Electrode Cell constant 0 9170 c Type of the temperature sen...

Page 10: ...an compensate for this influence This procedure determines a new cell constant for the system consisting of conductivity measuring cell measuring environment For more detailed information refer to sec...

Page 11: ...may be necessary to adapt the cell constant due to the influence of the measuring environ ment e g of boundary fields In chapter 6 WHAT TO DO IF of this operating manual you will find the correct valu...

Page 12: ...cell thoroughly Cleaning 5 3 Disposal We recommend to dispose of the conductivity measuring cell as elec tronic waste WARNING Contact with the sample can lead to danger to the user Depending on the t...

Page 13: ...Boundary field not maintained The conductivity measuring cell must be surrounded by a free space of at least 5 cm at the base and sides during measurement Otherwise the cell constant will change see...

Page 14: ...m without screwed cable gland Shaft diameter 40 mm Approx 660 g without sensor connection cable Sensor head PVC epoxy filler Electrodes housing of the tem perature sensor Graphite Shaft POM conductive...

Page 15: ...e of connection Cable with free stranded wire ends Terminal assignment WARNING All voltages must be protective low voltages or safety extra low voltages without hazard of contact according to E 61010...

Page 16: ...ve 97 23 EC pressure equipment directive 7 5 Characteristic data on delivery Temperature measurement Cell constant Measuring medium 0 C 40 C Storage transport 5 C 65 C storage in air recommended Max a...

Page 17: ......

Page 18: ...buildings factories and farms In more than 150 countries we have strong long standing relationships with customers who know us for our powerful combination of leading product brands and applications...

Reviews: