Maintenance, cleaning, replacement
StirrOx G
8
ba25310d07
08/2018
4.3
Storage
Always store the sensor in the calibration and storage vessel at a
temperature in the range 0 ... +50 °C (32 ... 122 °F). Make sure that the
sponge in the calibration and storage vessel is always moist.
Calibration and
storage vessel,
OxiCal
®
-ST
Moisten the sponge:
Screw out the insert (1).
Take out the sponge (2), wet it, then slightly squeeze it out.
Put in the sponge and screw the insert in the calibration and storage
vessel again.
Sta-ST stand
The Sta-ST stand (see chapter 8 W
EAR
PARTS
AND
ACCESSORIES
) enables
you to place the calibration and storage vessel with the sensor in an upright,
tilt resistant position on a table.
5
Maintenance, cleaning, replacement
5.1
General maintenance instructions
For your safety
Note the following safety instructions when handling electrolyte and cleaning
solutions:
CAUTION
The ELY/G electrolyte solution and RL-G cleaning solution irritate the
eyes and skin. Note the following points when dealing with these
solutions:
During working activities, always wear suitable protective gloves
and protective goggles/face shield.
If the solutions come into contact with the skin, rinse thoroughly with
water and immediately change contaminated clothing.
After each measurement series (before storing the sensor) clean
the sensor (exterior cleaning).