Technical data
17
ba25310d07
08/2018
Measurement
conditions
Measuring ranges at 20 °C
(68 °F)
0 ... 50 mg/l D.O.
0 ... 600 % D.O. saturation
0 ... 1250 mbar D.O. partial pressure
Temperature range
0 ... 50 °C (32 ... 122 °F)
Depth of immersion
(with stirring paddle)
min. 49 mm
max. 83 mm
Max. allowed overpressure
according to an immersion measurement up
to the max. immersion depth
Operating position
vertical with membrane cap downward
Approach flow
by integrated stirring system
Storage
conditions
Recommended storing
method
in the calibration and storage vessel,
OxiCal
®
-ST (moist)
Storage temperature
0 ... 50 °C (32 ... 122 °F)
Characteristics
when delivered
Zero signal
< 0.1 % of the saturation value
Response time of D.O.
measurement at 20 °C
(68 °F)
t
90
(90 % of the final value display after)< 10 s
t
95
(95 % of the final value display after)< 15 s
t
99
(99 % of the final value display after) < 45 s
Reproducibility
< 0.3 %
Own consumption of
oxygen at 20 °C (68 °F)
0.008
μ
g·h
-1
(mg/l)
-1
Drift
Approx. 3 % per month in the operating
condition
Response time of
temperature measurement
t
99
(99 % of the final value display after)< 60 s
Precision of temperature
measurement
± 0.2 K
Working life
min. 6 months with one electrolyte filling