4.4.6
Entering the salt content (salinity)
A salt content correction is required in the oxygen concentration mea-
surement of test samples with a salt content of more than 1 g/l. To do
this, you have to enter the salinity equivalent of the test sample (range
0.0 - 70.0) and to switch on the salinity correction.
Entering the salinity
Note
How to switch on the salt content correction is described on page 34.
1
Determine the salinity of the test sample.
2
Press the <CAL> key repeatedly until Sal appears on the dis-
play.
3
Enter the salt content with <
▲
> <
▼
>.
4
Switch to the measuring mode with <M>.
°