![wtw pH 197i Operating Manual Download Page 30](http://html1.mh-extra.com/html/wtw/ph-197i/ph-197i_operating-manual_1000648030.webp)
Technical data
pH 197i
30
Measuring ranges, resolutions, accuracies
* when measuring in a range of ± 2 pH around a calibration point
Analog output
(AK 323/S cable)
Automatic switchover when the recorder is connected by the cable,
AK 323/S.
pH output signal
-200 ... +1999 mV for the range
- 2.00 ... + 19.99
mV output signal
-1999 ... +1999 mV for the range
-1999 ... +1999 mV for the range
Accuracy
± 0.5 % of display value
Internal resistance
< 5 Ohm (current limited to max. 0.2 mA
output current)
Guidelines
and norms used
EMC
EC guideline 89/336/EEC
EN 61326-1:1998
EN 61000-3-2 A14:2000
EN 61000-3-3:1995
FCC Class A
Instrument safety
EC guideline 73/23/EEC
EN 61010-1 A2:1995
Climatic class
VDI/VDE 3540
IP protection
EN 60529:1991
Variable
Measuring
range
Resolution
Accuracy (± 1 Digit)
pH
- 2.00 ... + 19.99
0.01
± 0.01 *
U [mV]
- 199.9 ... + 199.9
- 1999 ... + 1999
0.1
1
± 0.5
± 1
T [°C]
- 5.0 ... + 105.0
0.1
± 0.1
Manual
temperature
input
Variable
Range
Increment
T
manual
[°C]
- 20 ... + 130
1