background image

Preparation instructions for P 700 IQ-solutions 

sure the funnel and the reactor walls are rinsed thoroughly. 

 

(Step does not apply for ST 0.0) 

  Then add the corresponding amount of hydrochloric acid (see Table 2-1) with the 

measuring cylinder. 

  Lastly, fill the reactor with the remaining demineralized water until the specified mass of 

water is reached. 

  Then switch on the agitator for at least 5 minutes. 

 

Prepared 

amount 

in L 

Sample weight of 

KH

2

PO

4

 

in g 

Sample weight of 

water amount 

in kg 

Addition to 

HCl 

in mL 

Total sample weight of 

water 

in g 

For ST 0.0 

10 

Approx. 8 

13 

9972 

20 

Approx. 17 

27 

19943 

30 

Approx. 25 

40 

29915 

40 

Approx. 30 

53 

39887 

50 

Approx. 40 

67 

49858 

60 

Approx. 50 

80 

59830 

For ST 1.6 

10 

0.0703 

Approx. 8 

13 

9972 

20 

0.1406 

Approx. 17 

27 

19943 

30 

0.2109 

Approx. 25 

40 

29915 

40 

0.2812 

Approx. 30 

53 

39887 

50 

0.3515 

Approx. 40 

67 

49858 

60 

0.4218 

Approx. 50 

80 

59830 

For ST 8.0 

10 

0.3515

 

Approx. 8 

13 

9972 

20 

0.7030

 

Approx. 17 

27 

19943 

30 

1.0545 

Approx. 25 

40 

29915 

40 

1.4059 

Approx. 30 

53 

39887 

50 

1.7574 

Approx. 40 

67 

49858 

60 

2.1090 

Approx. 50 

80 

59830 

For ST 20.0 

10 

0.8787 

Approx. 8 

13 

9972 

20 

1.7575 

Approx. 17 

27 

19943 

30 

2.6361 

Approx. 25 

40 

29915 

40 

3.5149 

Approx. 30 

53 

39887 

50 

4.3936 

Approx. 40 

67 

49858 

60 

5.2724 

Approx. 50 

80 

59830 

Table 2-1: Sample weights depending on the prepared amount (sample weight of water for 

ξ = 18 °C) and standard solution 

 

Testing 

The quality of the prepared phosphate phosphor standard solutions should be checked using 
a suitable measuring system

1

. The measuring system and the calibration standard must be 

prepared for this as per Chapter 2.1. Calibrating with traceable standards ensures the 
traceability to reference material by the NIST/PTB. 
 

  The standard solution is checked using a suitable measuring system 

  The presence of chloride must also be checked (qualitative testing for the use of 

added hydrochloric acid). 
 

 

                                                           

1

 

Accuracy ≤ 1 % (deviation of the measuring system) 

Summary of Contents for P 700 IQ

Page 1: ...ba77185e01 11 2016 Preparation instructions P 700 IQ Standards reagents cleaning solutions Orthophosphate Analyzer for measuring phosphate phosphorus...

Page 2: ...Preparation instructions for P 700 IQ solutions 2 Copyright 2016 by Xylem Analytics Germany GmbH...

Page 3: ...cations 6 1 4 Storage shelf life transport 6 2 Preparation instructions for standard solutions ST 0 0 ST 1 6 ST 8 0 ST 20 0 7 2 1 Preparation of the measuring system for PO4 P measurement calibration...

Page 4: ...Preparation instructions for P 700 IQ solutions 4...

Page 5: ...ust possess at least one certificate based on ISO 9001 For the preparation of standard solutions the producer must be able to furnish a certificate of traceability to secondary material in accordance...

Page 6: ...nteed by the customers producers 1 3 Chemicals and purity specifications Always use all chemicals in the specified purity as a rule p A Impure chemicals can falsify the measured result of the P 700 IQ...

Page 7: ...balance 0 1 g with summing function Reactor with agitator Measuring beaker 5 L Funnel Measuring cylinder 100 mL Beaker 50 mL pH indicator rods e g pH 0 6 Preparations Check reactors agitators vessels...

Page 8: ...18 Approx 50 80 59830 For ST 8 0 10 0 3515 Approx 8 13 9972 20 0 7030 Approx 17 27 19943 30 1 0545 Approx 25 40 29915 40 1 4059 Approx 30 53 39887 50 1 7574 Approx 40 67 49858 60 2 1090 Approx 50 80 5...

Page 9: ...urther testing Testing for chloride Qualitative testing only e g photometric or titrimetric Testing the pH value A representative sample of the test solution is taken for the pH test The sample must n...

Page 10: ...35 C is not a problem In the worst case irreversible damage from precipitates or even frost damage can occur below 5 C For subsequent use depending on the weather temperatures of 15 C 45 C are obtain...

Page 11: ...oughly Concentration of the solution to be tested in mgl l PO4 3 P Volumetric flask volume to be used in ml Volume of stock solution to be pipetted in ml Addition of fuming hydrochloric acid in ml 1 6...

Page 12: ...ed above for cleanliness clean if necessary Fill the measuring beaker with demineralized water and then empty again Tare the still wet measuring beaker using the precision balance Production to proces...

Page 13: ...ulfilled may the test solution be re utilized for production Otherwise this formulation must be discarded Filling After passing the test at least 2500 mL of the standard solution is poured into the ap...

Page 14: ...tion instructions for P 700 IQ solutions 14 Disposal According to current national provisions Safety instructions The current national provisions must be taken into account when handling hazardous sub...

Page 15: ...ssels and other aids and appliances mentioned above for cleanliness clean if necessary The measuring beaker is filled with demineralized water from the conduit system and then emptied again The precis...

Page 16: ...ded Filling After passing the test at least 1000 mL of the cleaning solution is poured into the approved containers and sealed tight with the associated closure containers as per Chapter 5 1 The conta...

Page 17: ...Preparation instructions for P 700 IQ solutions 17 Safety instructions The current national provisions must be taken into account when handling hazardous substances...

Page 18: ...s Self adhesive 105x148 mm or similar printable labels must be used The color dot for coding the hose connections must be taken into account The possibility for color printing is thus mandatory The fi...

Page 19: ...editing Hersteller Producer with address tel e mail Logo Brand Best nr Order no Reagent solution RE 2 5 WTW article no 821999 File template 28207_re2 5 tif Fields for editing Hersteller Producer with...

Reviews: