Preparation instructions for P 700 IQ-solutions 

16 

Prepared 

amount 

in L 

Amount of water 

in kg 

Sample weight 

of NaOH 

in g 

Sample weight  

Titriplex III 

in g 

Total sample 

weight of water 

in g 

10 

10 

400 

390 

9255 

20 

20 

800 

780 

18510 

30 

20 

1200 

1170 

27726 

40 

25 

1600 

1560 

36968 

50 

30 

2000 

1950 

46210 

60 

40 

2400 

2340 

55452 

Table 4-1: Sample weights depending on the prepared amount (sample weight of water for 

ξ = 18 °C) 

 
 

Testing 

The cleaning solution is subjected to a test by conductivity measurement. 
The conductivity measuring instrument and the conductivity cell must be calibrated or 
checked in accordance with the operating manual of the measuring instrument if necessary. 
The following settings must be made on the conductivity measuring instrument: 

  T

Ref

 = 25 °C 

  Temperature compensation "nLF" 

  Cell constant in accordance with conductivity cell 

If the measured conductivity is in the range of 275 mS/cm ± 20 mS/cm, the cleaning solution 
has been properly prepared. 
 
Using the test solution: 
Only if the above criteria are fulfilled may the prepared cleaning solution be re-utilized for 
production. Otherwise this formulation must be discarded. 
 

Filling 

After passing the test, at least 1000 mL of the cleaning solution is poured into the approved 
containers and sealed tight with the associated closure (containers as per Chapter 5.1).  
The containers are provided with the corresponding labels (see Chapter 5.2) 
 

Shelf life 

A shelf life as per Table 4-2 can be warranted for the cleaning solution prepared in this way. 
It is printed with the label (see Chapter 5.2) on the containers in MM.YY format. 
 

Solution 

Shelf life 

CL 1.0 

24 months 

Table 4-2: Shelf lives for prepared cleaning solution 

 

Storage: 

A shelf life for the cleaning solution as of production according to Table 4-2 is specified, from 
which the shelf life data on the label of the respective container is derived. 
For storage prior to use (transport, distribution center), a storage temperature of +5 °C…25 
°C is recommended. Brief heating during this time to  35 °C is not a problem. In the worst 
case, irreversible damage from precipitates or even frost damage can occur below 5 °C. 
For subsequent use, depending on the weather, temperatures of +15 °C…+45 °C are 
obtained in the P 700 IQ which is why the labeled shelf life date continues to be valid without 
restriction. 
Storage times must be limited to max. 6 months in order to safeguard corresponding usability 
for the user. 
 

Disposal 

According to current national provisions 
 

 

 

Summary of Contents for P 700 IQ

Page 1: ...ba77185e01 11 2016 Preparation instructions P 700 IQ Standards reagents cleaning solutions Orthophosphate Analyzer for measuring phosphate phosphorus...

Page 2: ...Preparation instructions for P 700 IQ solutions 2 Copyright 2016 by Xylem Analytics Germany GmbH...

Page 3: ...cations 6 1 4 Storage shelf life transport 6 2 Preparation instructions for standard solutions ST 0 0 ST 1 6 ST 8 0 ST 20 0 7 2 1 Preparation of the measuring system for PO4 P measurement calibration...

Page 4: ...Preparation instructions for P 700 IQ solutions 4...

Page 5: ...ust possess at least one certificate based on ISO 9001 For the preparation of standard solutions the producer must be able to furnish a certificate of traceability to secondary material in accordance...

Page 6: ...nteed by the customers producers 1 3 Chemicals and purity specifications Always use all chemicals in the specified purity as a rule p A Impure chemicals can falsify the measured result of the P 700 IQ...

Page 7: ...balance 0 1 g with summing function Reactor with agitator Measuring beaker 5 L Funnel Measuring cylinder 100 mL Beaker 50 mL pH indicator rods e g pH 0 6 Preparations Check reactors agitators vessels...

Page 8: ...18 Approx 50 80 59830 For ST 8 0 10 0 3515 Approx 8 13 9972 20 0 7030 Approx 17 27 19943 30 1 0545 Approx 25 40 29915 40 1 4059 Approx 30 53 39887 50 1 7574 Approx 40 67 49858 60 2 1090 Approx 50 80 5...

Page 9: ...urther testing Testing for chloride Qualitative testing only e g photometric or titrimetric Testing the pH value A representative sample of the test solution is taken for the pH test The sample must n...

Page 10: ...35 C is not a problem In the worst case irreversible damage from precipitates or even frost damage can occur below 5 C For subsequent use depending on the weather temperatures of 15 C 45 C are obtain...

Page 11: ...oughly Concentration of the solution to be tested in mgl l PO4 3 P Volumetric flask volume to be used in ml Volume of stock solution to be pipetted in ml Addition of fuming hydrochloric acid in ml 1 6...

Page 12: ...ed above for cleanliness clean if necessary Fill the measuring beaker with demineralized water and then empty again Tare the still wet measuring beaker using the precision balance Production to proces...

Page 13: ...ulfilled may the test solution be re utilized for production Otherwise this formulation must be discarded Filling After passing the test at least 2500 mL of the standard solution is poured into the ap...

Page 14: ...tion instructions for P 700 IQ solutions 14 Disposal According to current national provisions Safety instructions The current national provisions must be taken into account when handling hazardous sub...

Page 15: ...ssels and other aids and appliances mentioned above for cleanliness clean if necessary The measuring beaker is filled with demineralized water from the conduit system and then emptied again The precis...

Page 16: ...ded Filling After passing the test at least 1000 mL of the cleaning solution is poured into the approved containers and sealed tight with the associated closure containers as per Chapter 5 1 The conta...

Page 17: ...Preparation instructions for P 700 IQ solutions 17 Safety instructions The current national provisions must be taken into account when handling hazardous substances...

Page 18: ...s Self adhesive 105x148 mm or similar printable labels must be used The color dot for coding the hose connections must be taken into account The possibility for color printing is thus mandatory The fi...

Page 19: ...editing Hersteller Producer with address tel e mail Logo Brand Best nr Order no Reagent solution RE 2 5 WTW article no 821999 File template 28207_re2 5 tif Fields for editing Hersteller Producer with...

Reviews: