8.4
Options for data transmission to a PC
Via the USB-B interface you can transmit data to a PC. The following table
shows which data are transmitted to the interface in which way:
8.5
MultiLab Importer
With the aid of the MultiLab Importer software, you can record and evaluate
measuring data with a PC.
Data
Control
Operation / description
Current
measured
values of all
connected
sensors
Manual
With
<F2>
/
[USB output]
.
Simultaneously with every manual
data storage process (see section 7.1
M
Automatic, at
intervals
With
<F2__>
/
[USB output]
.
Then you can set the transmission
interval.
Simultaneously with every automatic
data storage process (see section 7.2
A
Stored mea-
sured values
Manual
Displayed dataset with
<F2>
[USB out-
put]
after calling up from the data stor-
age.
All datasets with the
Output to RS232/
USB
function.
calibration
records
Manual
Calibration record with
<F2>
/
[USB out-
put]
(see section 5.3.6 C
Automatic
At the end of a calibration procedure.
The following rule applies: With the exception of the menus, shortly
pressing the
<F2>
/
[USB output]
key generally outputs the display
contents to the interface (displayed measured values, measuring
datasets, calibration records).
More detailed information can be found in the MultiLab Importer
operating manual.