Multi 9310
What to do if...
ba75937e03
12/2011
89
No stable measured
value
Obviously incorrect
measured values
14.2 Dissolved oxygen
Error message,
OFL, UFL
The measured value is outside the measuring range.
Cause
Remedy
IDS pH sensor:
– Junction contaminated
– Clean junction
– Membrane contaminated
– Clean membrane
Test sample:
– pH value not stable
– Measure with air excluded if neces-
sary
– Temperature not stable
– Adjust temperature if necessary
IDS pH test sample:
– Conductivity too low
– Use suitable IDS pH sensor
– Temperature too high
– Use suitable IDS pH sensor
– Organic liquids
– Use suitable IDS pH sensor
Cause
Remedy
IDS pH sensor:
– IDS pH sensor unsuitable
– Use suitable IDS sensor
– Temperature difference
between buffer and test sample
too great
– Adjust temperature of buffer or sam-
ple solutions
– Measurement procedure not
suitable
– Follow special procedure
More information and instructions on cleaning and exchange of
sensors are given in the documentation of your sensor.